InterStim(TM) System, Model Numbers recallZ-1774-2019 · Medtronic Neuromodulation · initiated May 15, 2019

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Neuromodulation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for an unexpected increase in stimulation during InterStim programming with the A10 Clinician Application (on Medtronic's smart programmer).

FDA root cause category: Software design

What the company is doing

Medtronic sent an Urgent Medical Device Safety Notification letter dated May 2019, to US Physicians and European physician. The notifications were delivered by mail, personal delivery by Medtronic Representatives, fax, or equivalent method. A confirmation form will be used to document receipt and understanding of the notification, and a minimum three attempts will be made to obtain confirmation from non-responding physicians that the notification has been received and understood.

Recall details

Recall number
Z-1774-2019
Status
Open, Classified
Initiated by the company
May 15, 2019
Posted by the FDA
Jun 7, 2019
Recalling firm
Medtronic Neuromodulation
Quantity
13979 units
Product code
EZW
FDA submissions
P970004S268, P080025S163
Distribution
Worldwide Distribution - US Nationwide & PR, and Germany, Switzerland, Italy, Spain, France, UK, Norway, Denmark, Finland, Netherlands

Product description

InterStim(TM) System, Model Numbers: a) TH90G01 b) TH90GFA c) TH90G02 d) TH90G03 Product Usage: The Medtronic Model A510 Clinician application (app) is intended for use with the HH90 Handset and TM90 Communicator to program, adjust, and troubleshoot the Medtronic Models 3023 and 3058 InterStim neurostimulators for sacral neuromodulation therapy. The clinician uses the Clinician app to program settings for the patient. The A510 Clinician app, HH90 Handset, TM90 Communicator along with the A520 Patient app are only sold as a kit (TH90).

Code information (models, lots, serial numbers)

Model Numbers/UDI: a) TH90G01/00763000058005  b) TH90GFA/00763000187231 c) TH90G02/00763000192259, 00763000192266, 00763000192273,  00763000192280, 00763000192297, 00763000192303, 00763000192310 d) TH90G03/00763000192310  ALL LOT/SERIAL NUMBERS