Medical device recalls: June 2019
The FDA posted 57 medical device recall records in June 2019, from 17 recall events by 14 companies. 21 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Maquet Cardiovascular Us Sales, Llc · 2 events, 14 records
- GE Healthcare, LLC · 2 events, 11 records
- Cook Inc. · 2 events, 4 records, 3 Class I
- Teleflex Medical · 1 event, 8 records, 8 Class I
- OriGen Biomedical, Inc. · 1 event, 6 records, 6 Class I
- Medtronic Inc · 1 event, 5 records
- Vyaire Medical · 1 event, 2 records, 2 Class I
- Becton Dickinson & Company · 1 event, 1 record, 1 Class I
FDA root cause
- Device Design · 23
- Process control · 16
- Radiation Control for Health and Safety Act · 6
- Under Investigation by firm · 5
- Labeling mix-ups · 5
- Software design · 1
- Labeling False and Misleading · 1
Class I recalls (21)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1210-2019 · Becton Dickinson & Company · posted Jun 26, 2019 · Terminated
Leaking of the Smartsite Syringe Administration Set. May result in delay or interruption of infusion, under-infusion, contamination of the fluid path and HCP exposure to infusates.
- enFlow Disposable Cartridge, used for intravenous warming therapy with fluid and blood solutions.Class IZ-1385-2019 · Vyaire Medical · posted Jun 26, 2019 · Terminated
Testing has demonstrated aluminum elution from the enFlow Disposable Cartridge during fluid warming.
- Z-1386-2019 · Vyaire Medical · posted Jun 26, 2019 · Terminated
Testing has demonstrated aluminum elution from the enFlow Disposable Cartridge during fluid warming.
- Z-1759-2019 · Teleflex Medical · posted Jun 19, 2019 · Terminated
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Hudson RCI Sheridan:Class IZ-1760-2019 · Teleflex Medical · posted Jun 19, 2019 · Terminated
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Hudson RCI Sheridan:Class IZ-1761-2019 · Teleflex Medical · posted Jun 19, 2019 · Terminated
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Hudson RCI Sheridan:Class IZ-1762-2019 · Teleflex Medical · posted Jun 19, 2019 · Terminated
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Hudson RCI Sheridan:Class IZ-1763-2019 · Teleflex Medical · posted Jun 19, 2019 · Terminated
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Hudson RCI Sheridan:Class IZ-1764-2019 · Teleflex Medical · posted Jun 19, 2019 · Terminated
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Z-1765-2019 · Teleflex Medical · posted Jun 19, 2019 · Terminated
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Sheridan/HVT:Class IZ-1766-2019 · Teleflex Medical · posted Jun 19, 2019 · Terminated
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
- Z-1756-2019 · Cook Inc. · posted Jun 17, 2019 · Terminated
Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon…
- Z-1757-2019 · Cook Inc. · posted Jun 17, 2019 · Terminated
Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon…
- Z-1758-2019 · Cook Inc. · posted Jun 17, 2019 · Terminated
Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon…
- Z-1705-2019 · Edwards Lifesciences, LLC · posted Jun 11, 2019 · Terminated
Possibility of balloon rupture in intra-aortic occlusion devices. If the balloon bursts during cardiopulmonary bypass, the heart can fill and warm, the operative site may be obscured, and the device will need to be exchanged or operative strategy would…
- Z-1227-2019 · OriGen Biomedical, Inc. · posted Jun 4, 2019 · Terminated
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear…
- Z-1228-2019 · OriGen Biomedical, Inc. · posted Jun 4, 2019 · Terminated
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear…
- Z-1229-2019 · OriGen Biomedical, Inc. · posted Jun 4, 2019 · Terminated
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear…
- Z-1230-2019 · OriGen Biomedical, Inc. · posted Jun 4, 2019 · Terminated
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear…
- Z-1231-2019 · OriGen Biomedical, Inc. · posted Jun 4, 2019 · Terminated
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear…
- Z-1232-2019 · OriGen Biomedical, Inc. · posted Jun 4, 2019 · Terminated
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear…
Class II recalls (35)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1880-2019 · Medtronic Inc · posted Jun 26, 2019 · Terminated
Fifteen affected devices were distributed worldwide, which were mislabeled with an incorrect size. In all cases, the actual surgical heart valves are larger than what was indicated on the boxes and jar labels.
- Z-1881-2019 · Medtronic Inc · posted Jun 26, 2019 · Terminated
Fifteen affected devices were distributed worldwide, which were mislabeled with an incorrect size. In all cases, the actual surgical heart valves are larger than what was indicated on the boxes and jar labels.
- Z-1882-2019 · Medtronic Inc · posted Jun 26, 2019 · Terminated
Fifteen affected devices were distributed worldwide, which were mislabeled with an incorrect size. In all cases, the actual surgical heart valves are larger than what was indicated on the boxes and jar labels.
- Z-1883-2019 · Medtronic Inc · posted Jun 26, 2019 · Terminated
Fifteen affected devices were distributed worldwide, which were mislabeled with an incorrect size. In all cases, the actual surgical heart valves are larger than what was indicated on the boxes and jar labels.
- Z-1884-2019 · Medtronic Inc · posted Jun 26, 2019 · Terminated
Fifteen affected devices were distributed worldwide, which were mislabeled with an incorrect size. In all cases, the actual surgical heart valves are larger than what was indicated on the boxes and jar labels.
- Z-1852-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1853-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1854-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1855-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1856-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1857-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1858-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1859-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1860-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1861-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1862-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1863-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Z-1864-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 21, 2019 · Terminated
Potential that the coating of the light head forks on the VOLISTA Surgical Lights may crack, posing the risk of particles falling into the surgical field during surgery that may result in patient infection
- Discovery PET/CT 610 ↗Class IIZ-1745-2019 · GE Healthcare, LLC · posted Jun 19, 2019 · Open, Classified
GE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
- Discovery PET/CT 710Class IIZ-1746-2019 · GE Healthcare, LLC · posted Jun 19, 2019 · Open, Classified
GE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
- Optima PET/CT 560, 560FX ↗Class IIZ-1747-2019 · GE Healthcare, LLC · posted Jun 19, 2019 · Open, Classified
GE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
- Discovery IQClass IIZ-1748-2019 · GE Healthcare, LLC · posted Jun 19, 2019 · Open, Classified
GE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
- Discovery MlClass IIZ-1749-2019 · GE Healthcare, LLC · posted Jun 19, 2019 · Open, Classified
GE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
- Discovery Ml DRClass IIZ-1750-2019 · GE Healthcare, LLC · posted Jun 19, 2019 · Open, Classified
GE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
- Z-1844-2019 · Cook Inc. · posted Jun 19, 2019 · Terminated
may contain a damaged gray safety lock knob, which could potentially result in difficulty or an inability to fully deploy the graft via the standard or troubleshooting method provided in the Instructions for Use (IFU). Potential adverse events that may…
- Z-1957-2019 · Richard Wolf Medical Instruments Corp. · posted Jun 13, 2019 · Terminated
E-Line Cutting Electrode package 8416.0305 contains E-Line Hook Electrode 8416.09 instead of E-Line Cutting Electrode 8416.03.
- Z-1751-2019 · GE Healthcare, LLC · posted Jun 10, 2019 · Terminated
There has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.
- Proteus XR/aClass IIZ-1752-2019 · GE Healthcare, LLC · posted Jun 10, 2019 · Terminated
There has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.
- Prestilix 1600X ↗Class IIZ-1753-2019 · GE Healthcare, LLC · posted Jun 10, 2019 · Terminated
There has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.
- Precision 500D ↗Class IIZ-1754-2019 · GE Healthcare, LLC · posted Jun 10, 2019 · Terminated
There has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.
- Silhouette VR ↗Class IIZ-1755-2019 · GE Healthcare, LLC · posted Jun 10, 2019 · Terminated
There has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.
- Z-1774-2019 · Medtronic Neuromodulation · posted Jun 7, 2019 · Open, Classified
There is a potential for an unexpected increase in stimulation during InterStim programming with the A10 Clinician Application (on Medtronic's smart programmer).
- Z-1729-2019 · Leica Microsystems, Inc. · posted Jun 4, 2019 · Terminated
The current labelling does not include any specific warning about usage of flammable freezing sprays. Such flammable freezing sprays can potentially ignite and can cause serious injuries.
- Z-1734-2019 · Maquet Cardiovascular Us Sales, Llc · posted Jun 4, 2019 · Terminated
User may suffer an electric shock when coming in contact with the exposed electronic components
- Z-1708-2019 · Stryker Neurovascular · posted Jun 2, 2019 · Terminated
A design related issue can contribute to the device exhibiting brittle behavior. This behavior may result in fractures of the catheter shaft.
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1843-2019 · Cytocell Ltd. · posted Jun 19, 2019 · Terminated
There is an error in the chromomap on the package insert.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.