Stryker Neurovascular, AXS Infinity LS Plus Long Sheath, Rx… recallZ-1708-2019 · Stryker Neurovascular · initiated Apr 18, 2019
Class II Terminated All recalls posted in June 2019
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Neurovascular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A design related issue can contribute to the device exhibiting brittle behavior. This behavior may result in fractures of the catheter shaft.
FDA root cause category: Device Design
What the company is doing
On 04/18/19, Urgent Medical Device Voluntary Recall-Removal notices were sent to customers via overnight delivery. Customers were asked to do the following: 1. Immediately check your internal inventory for affected devices. 2. Segregate the affected units in a secure location for return to firm. 3. Circulate this Field Safety Notice internally to all interested/affected parties. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. Inform recalling firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so that the recalling firm can inform the recipients appropriately. 6. Please inform the recalling firm of any adverse events concerning the use of the subject devices. 7. Complete the attached customer response form. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. 8. Email the completed form to [email protected] 9. Product Return Information will be provided to you by your designated Territory Manager or Sales Representative. In addition, If you have loaned or sold any of the units listed, please, forward a copy of this notice to the new users and advise the recalling firm of their new location.
Recall details
- Recall number
- Z-1708-2019
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 18, 2019
- Posted by the FDA
- Jun 2, 2019
- Recalling firm
- Stryker Neurovascular
- Quantity
- 2435
- Product code
- DQY
- FDA submissions
- K172468
- Distribution
- U.S Nationwide Distribution - PA, VA, TX, CA, FL, AR, TN, MI, NY, IL, NC, KY, WV, OH, MO, AZ, GA, WA, VT, DC, SC, NJ, MN, OK, MD, AL, LA, MA, DE, IN, OR, RI, SD, MS, WI, ID, UT, CO, IA, KS, CT, NE, NM
Product description
Stryker Neurovascular, AXS Infinity LS Plus Long Sheath, Rx Only, Sterile EO, USA Customer Service 855-91 NEURO (916-3876): a) REF: INC-11196-70, 70cm, b) REF: INC-11196-80, 80cm, c) REF: INC-11196-90, 90cm Product Usage: Indications for Use: The AXS Infinity LS Plus Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. Device Description: The AXS Infinity LS Plus Long Sheath is a single lumen, flexible, variable stiffness long sheath with an 0.091 inch inner diameter. It has a radiopaque marker band on the distal end and a Luer hub at the proximal end. The AXS Infinity LS Plus Long Sheath shaft has a 10 cm lubricious coating at the distal end to reduce friction during use. The inner lumen of the AXS Infinity LS Plus Long Sheath is compatible with 6F or smaller catheters. Each package includes one AXS Infinity LS Plus Long Sheath (INC- 11196-70, INC-11196-80, or INC-11196-90), one Dilator, and one hemostasis valve. Dimensions of the AXS Infinity LS Plus Long Sheath are included on the individual device label.
Code information (models, lots, serial numbers)
All lots
Other devices under the same FDA submission
3 devices cleared under K172468 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- AXS INFINITY LS PLUS INC-11196-90Stryker Corporation · Catheter, percutaneous26 reports
- AXS INFINITY LS PLUS INC-11196-70Stryker Corporation · Catheter, percutaneous2 reports
- AXS INFINITY LS PLUS INC-11196-80Stryker Corporation · Catheter, percutaneous2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page