Skip to content

BD MicrotainerTubes with K2E (K2EDTA) recallZ-0057-2020 · Becton Dickinson & Company · initiated May 13, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results.

FDA root cause category: Process control

What the company is doing

Becton Dickinson (BD) notified consignees affected by letter via UPS and email dated May 13,2019. Letter describes reason for recall, health risk and action to take: 1. Immediately review your inventory for the specific catalog and lot numbers listed below, 2. Destroy all product subject to the recall following your institution's process for destruction, 3. Share this Urgent Medical Device Recall notification with all users of the product to ensure that they are also aware of this recall, and 4. Complete the attached Customer Response/Certificate of Destruction Forms and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement. Distributors provide their customer list to BD. BD PAS will notify these customers via mail. BD Contact US Contact Information Customer/Technical Support 888-237-2762 OPT 3, OPT 2; Monday Friday 8:00am - 5:00pm (CT).

Recall details

Recall number
Z-0057-2020
Status
Terminated
Initiated by the company
May 13, 2019
Recalling firm
Becton Dickinson & Company
Product code
JKA
FDA submissions
K002456, K991702
Distribution
Nationwide Foreign: AR AU BE BN BR CA CL CN CO CR CW EC GT GY HK HN ID IN JP KR MX MY NI NZ PA PE PG PH PK SG SR SV TH TW UY VN

Product description

BD MicrotainerTubes with K2E (K2EDTA) Catalog Number: 365974

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

Lot Number/UDI: 8101695 (01)30382903659747 (17)190930(10)8081695(30)50 (17)190930(10)8101695(30)200(01)50382903659741   8101696 (01)30382903659747 (17)190930(10)8081696(30)50 (17)190930(10)8101696(30)200(01)50382903659741   8101699 (01)30382903659747 (17)190930(10)8101699(30)50 (17)190930(10)8101699(30)200(01)50382903659741   8101701 (01)30382903659747 (17)190930(10)8101701(30)50 (17)190930(10)8101701(30)200(01)50382903659741    8106835 (01)30382903659747 (17)190930(10)8106835(30)50 (17)190930(10)8106835(30)200(01)50382903659741   8106837 (01)30382903659747 (17)190930(10)8106837(30)50 (17)190930(10)8106837(30)200(01)50382903659741   8106842 (01)30382903659747 (17)190930(10)8106842(30)50 (17)190930(10)8106842(30)200(01)50382903659741   8106847 (01)30382903659747 (17)190930(10)8106847(30)50 (17)190930(10)8106847(30)200(01)50382903659741   8106850 (01)30382903659747 (17)190930(10)8106850(30)50 (17)190930(10)8106850(30)200(01)50382903659741   8120851 (01)30382903659747 (17)191031(10)8120850(30)50 (17)191031(10)8120850(30)200(01)50382903659741     8121979 (01)30382903659747 (17)191031(10)8121979(30)50 (17)191031(10)8121979(30)200(01)50382903659741   8121981 (01)30382903659747 (17)191031(10)8121981(30)50 (17)191031(10)8121981(30)200(01)50382903659741   8124783 (01)30382903659747 (17)191031(10)8124783(30)50 (17)191031(10)8124783(30)200(01)50382903659741   8124880 (01)30382903659747 (17)191031(10)8124880(30)50 (17)191031(10)8124880(30)200(01)50382903659741   8127770 (01)30382903659747 (17)191031(10)8127770(30)50 (17)191031(10)8127770(30)200(01)50382903659741   8128575 (01)30382903659747 (17)191031(10)8128575(30)50 (17)191031(10)8128575(30)200(01)50382903659741   8129642 (01)30382903659747 (17)191031(10)8129642(30)50 (17)191031(10)8129642(30)200(01)50382903659741   8129645 (01)30382903659747 (17)191031(10)8129645(30)50 (17)191031(10)8129645(30)200(01)50382903659741   8197705 (01)30382903659747 (17)191231(10)8197705(30)50 (17)191231(10)8197705(30)200(01)50382903659741   8197970 (01)30382903659747 (17)191231(10)8197970(30)50 (17)191231(10)8197970(30)200(01)50382903659741   8198921 (01)30382903659747 (17)191231(10)8198921(30)50 (17)191231(10)8198921(30)200(01)50382903659741   8198927 (01)30382903659747 (17)191231(10)8198927(30)50 (17)191231(10)8198927(30)200(01)50382903659741   8243545 (01)30382903659747 (17)200229(10)8243545(30)50 (17)200229(10)8243545(30)200(01)50382903659741   8247628 (01)30382903659747 (17)200229(10)8247628(30)50 (17)200229(10)8247628(30)200(01)50382903659741   8248705 (01)30382903659747 (17)200229(10)8248705(30)50 (17)200229(10)8248705(30)200(01)50382903659741   8249986 (01)30382903659747 (17)200229(10)8249986(30)50 (17)200229(10)8249986(30)200(01)50382903659741   8288904 (01)30382903659747 (17)200331(10)8288904(30)50 (17)200331(10)8288904(30)200(01)50382903659741   8290928 (01)30382903659747 (17)200331(10)8290928(30)50 (17)200331(10)8290928(30)200(01)50382903659741"

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

7 devices cleared under K002456, K991702 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 82851).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page