HUDSON RCI recallZ-1912-2019 · Teleflex Medical · initiated May 10, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Teleflex Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Reported complaints of cracks being observed, prior to use, on swivel wye adaptors of the Neonatal ConchaSmart Breathing Circuit.

FDA root cause category: Process control

What the company is doing

The firm initiated the recall by email on May 10, 2019. The recall notice requested the return of the product. The letter identified the affected product, problem and actions to be taken. Customers were instructed to 1. If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to [email protected]. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. For questions contact Customer Service at 1-866-396-2111.

Recall details

Recall number
Z-1912-2019
Status
Terminated
Initiated by the company
May 10, 2019
Posted by the FDA
Jul 15, 2019
Recalling firm
Teleflex Medical
Quantity
20 units
Product code
BZE
FDA submissions
K173280
Distribution
US Nationwide Distribution in the states of AL, TX

Product description

Teleflex HUDSON RDI NEONATAL ConchaSmart Breathing Circuit with Dual Heated Limb and ConchaSmart Column: a) REF 870-07KIT b) REF 870-09KIT Product Usage: The Neonatal ConchaSmart breathing circuits are intended to deliver humidified breathing gases for administration to an infant/neonate patient. These gases may be delivered by nasal prongs using the CareFusion Infant Flow system and the single limb circuit, or through bypassing upper airways, for example through the use of an endotracheal tube with the dual limb configuration.

Code information (models, lots, serial numbers)

Lot/ Batch Number GTIN a) 74L1802045 14026704611154 b) 74L1802044 14026704611161

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates