Medical device recalls: July 2019
The FDA posted 50 medical device recall records in July 2019, from 17 recall events by 15 companies. 13 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- GE Healthcare, LLC · 3 events, 6 records, 2 Class I
- Burlington Medical, LLC · 1 event, 24 records
- Datascope Corp. · 1 event, 6 records, 6 Class I
- Roche Diagnostics Corporation · 1 event, 2 records
- Zevex Incorporated (dba MOOG Medical Devices Group) · 1 event, 2 records
- Becton Dickinson & Company · 1 event, 1 record, 1 Class I
- CareFusion 303, Inc. · 1 event, 1 record, 1 Class I
- Centurion Medical Products Corporation · 1 event, 1 record, 1 Class I
FDA root cause
- Under Investigation by firm · 24
- Device Design · 8
- Use error · 6
- Nonconforming Material/Component · 5
- Software design · 2
- Process control · 2
- Error in labeling · 1
- Employee error · 1
Class I recalls (13)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1956-2019 · Centurion Medical Products Corporation · posted Jul 30, 2019 · Terminated
Centurion is conducting a sub-recall for Teleflex Sheridan branded Endotracheal Tracheal Tubes due to increased incidence of Sheridan connector becoming disconnected from Endotracheal Tube.
- Alaris Pump, Model 8100, with bezel mechanical assemblies manufactured with FR-110 plastic resin.Class IZ-1768-2019 · CareFusion 303, Inc. · posted Jul 16, 2019 · Terminated
The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e. external cracking). The…
- Z-1939-2019 · Datascope Corp. · posted Jul 16, 2019 · Open, Classified
Battery short-run times and unanticipated stoppage may be due to user not performing battery maintenance and handling as needed. Inadequate labeling is being targeted for this recall, and design updates will be handled in a subsequent recall.
- Z-1940-2019 · Datascope Corp. · posted Jul 16, 2019 · Open, Classified
If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being provided.
- Z-1941-2019 · Datascope Corp. · posted Jul 16, 2019 · Open, Classified
If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being offered.
- Z-1942-2019 · Datascope Corp. · posted Jul 16, 2019 · Open, Classified
If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being provided.
- Z-1943-2019 · Datascope Corp. · posted Jul 16, 2019 · Open, Classified
If battery maintenance is not performed per instructions, the battery may provide less than the minimum run time of operating power.
- Z-1944-2019 · Datascope Corp. · posted Jul 16, 2019 · Open, Classified
If battery maintenance is not performed per instructions, the battery may provide less than the minimum run time of operating power.
- Z-1912-2019 · Teleflex Medical · posted Jul 15, 2019 · Terminated
Reported complaints of cracks being observed, prior to use, on swivel wye adaptors of the Neonatal ConchaSmart Breathing Circuit.
- Z-1770-2019 · Becton Dickinson & Company · posted Jul 11, 2019 · Terminated
An incomplete occlusion can be caused by a variation in the wall thickness of the pumping segment of the affected infusion sets used with the. The issue has the potential to lead to unintended delivery of medication when the pump module is not in running…
- Z-1846-2019 · GE Healthcare, LLC · posted Jul 11, 2019 · Terminated
The bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle.
- Panda Infant Radiant WarmerClass IZ-1847-2019 · GE Healthcare, LLC · posted Jul 11, 2019 · Terminated
The bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle.
- Z-1874-2019 · Hamilton Medical AG · posted Jul 11, 2019 · Terminated
New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator…
Class II recalls (35)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2107-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2108-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Frontal Support Belt Apron, Product Number F-15, Part Numbers REF15*F, REF15*M, REF15B, REF15XLM ↗Class IIZ-2109-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2110-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2111-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2112-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2113-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2114-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2115-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2116-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2117-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2118-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2119-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2120-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2121-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2122-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Sleeve, Product Number Sleeve, Part Numbers RESL, SL 3 L, SL 3 M, SL 3 S, L 3 X, SL 3 XS, UESL ↗Class IIZ-2123-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2124-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2125-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2126-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2127-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2128-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2129-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2130-2019 · Burlington Medical, LLC · posted Jul 30, 2019 · Terminated
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables…
- Z-2053-2019 · Zevex Incorporated (dba MOOG Medical Devices Group) · posted Jul 23, 2019 · Terminated
Administration sets leaked at the filter.
- Z-2054-2019 · Zevex Incorporated (dba MOOG Medical Devices Group) · posted Jul 23, 2019 · Terminated
Administration sets leaked at the filter.
- Z-2055-2019 · InfuTronix LLC · posted Jul 23, 2019 · Terminated
Potential for tube leakage
- Z-1984-2019 · Medline Industries, Inc. · posted Jul 18, 2019 · Open, Classified
The Anti-Fog Solution may not defog properly due to a degradation in the chemical properties.
- Z-1980-2019 · Elekta, Inc. · posted Jul 16, 2019 · Terminated
The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.
- Z-1964-2019 · GE Healthcare, LLC · posted Jul 12, 2019 · Terminated
Potential for a transducer mis-alignment in certain transvaginal probes.
- Repair RIC5-9A-RS System, imaging, pulsed doppler, ultrasonic and Refurbish RIC-5-9-D REALTIME 4DClass IIZ-1965-2019 · GE Healthcare, LLC · posted Jul 12, 2019 · Terminated
Potential for a transducer mis-alignment in certain transvaginal probes.
- Z-1721-2019 · Kerr/Pentron, DBA Kerr Corporation and Pentron Clinical · posted Jul 5, 2019 · Terminated
Immersion time stated on the bottle is 45 minutes while the product instructions say 90 minutes.
- Z-1923-2019 · Philips North America, LLC · posted Jul 5, 2019 · Terminated
The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring and failure or delay in the delivery of a shock or…
- Centricity Universal Viewer 6.0Class IIZ-1910-2019 · GE Healthcare, LLC · posted Jul 1, 2019 · Terminated
There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive.
- Centricity PACS FoundationClass IIZ-1911-2019 · GE Healthcare, LLC · posted Jul 1, 2019 · Terminated
There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive.
Class III recalls (2)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1927-2019 · Roche Diagnostics Corporation · posted Jul 3, 2019 · Terminated
There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.
- Z-1928-2019 · Roche Diagnostics Corporation · posted Jul 3, 2019 · Terminated
There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.