Babi.Plus 12.5 cm H2O Pressure Relief Manifold Catalog #… recallZ-0009-2020 · Teleflex-Morrisville · initiated Jun 14, 2019
Class I Terminated All recalls posted in November 2019
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Teleflex-Morrisville or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Device vented gas below the stated pressure.
FDA root cause category: Device Design
What the company is doing
On June 14 the firm sent communication with the following instructions: 1. Immediately discontinue distribution and quarantine any products with the product codes and lot numbers listed above. 2. Using the provided customer letter and recall acknowledgement form templates, communicate this recall to any of your customers who have received product included within the scope of the recall. 3. Have the customers return any affected product to you, together with a completed Recall Acknowledgement Form, for consolidation and return to Teleflex Medical. In the event that an alternative approach is needed, contact Teleflex Medical Customer Service for more information at 1-866-246-6990. 4. To return recalled products to Teleflex Medical, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of product to Teleflex Medical. 5. Once you have completed returning all of the recalled products from your own inventory, and collecting and consolidating all of the recalled products from your customers, please check the box on the enclosed Recall Acknowledgment Form that indicates that you have completed the recall and fax it to 1-855-419-8507, Attn: Customer Service, or email it to [email protected]. This will allow us to document completion of the recall. 6. If you and your customers have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. This will allow us to document your receipt of this letter.
Recall details
- Recall number
- Z-0009-2020
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Jun 14, 2019
- Posted by the FDA
- Nov 7, 2019
- Recalling firm
- Teleflex-Morrisville
- Quantity
- 9,312 each
- Product code
- BZD
- FDA submissions
- K110383
- Distribution
- US nationwide distribution. AL, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TX, VA, WA and WY
Product description
Babi.Plus 12.5 cm H2O Pressure Relief Manifold Catalog # 2691 for use with ventilators.
Code information (models, lots, serial numbers)
Catalog # 2691 SKU #: XTF0046 Lot #'s: 180806, 180910, 181029, 181105, 181204, 190225 and 190327
Other devices under the same FDA submission
1 device cleared under K110383 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Babi.Plus BC23-HDGalemed Corporation · Ventilator, Non-Continuous (Respirator)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page