Medical device recalls: November 2019
The FDA posted 26 medical device recall records in November 2019, from 9 recall events by 9 companies. 9 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Boston Scientific Corporation · 1 event, 13 records
- Philips Medical Systems (Cleveland) Inc · 1 event, 5 records, 5 Class I
- Medtronic Inc. · 1 event, 2 records, 2 Class I
- Abbott Gmbh & Co. KG · 1 event, 1 record
- Implant Direct Sybron Manufacturing, LLC · 1 event, 1 record
- Medtech Sas · 1 event, 1 record, 1 Class I
- Philips North America, LLC · 1 event, 1 record
- Spineart SA · 1 event, 1 record
FDA root cause
- Incorrect or no expiration date · 13
- Nonconforming Material/Component · 5
- Software design · 2
- Software in the Use Environment · 1
- Process control · 1
- Packaging · 1
- Material/Component Contamination · 1
- Employee error · 1
Class I recalls (9)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0009-2020 · Teleflex-Morrisville · posted Nov 7, 2019 · Terminated
Device vented gas below the stated pressure.
- Z-0177-2020 · Medtronic Inc. · posted Nov 4, 2019 · Terminated
There is a potential security vulnerability related to the use of the remote controller accessories with the insulin pumps.
- Z-0178-2020 · Medtronic Inc. · posted Nov 4, 2019 · Terminated
There is a potential security vulnerability related to the use of the remote controller accessories with the insulin pumps.
- Z-0172-2020 · Medtech Sas · posted Nov 1, 2019 · Terminated
The instrument holder may be sent on a trajectory that is not within the intended target. If it is not corrected, the associated device may be placed incorrectly.
- Z-0213-2020 · Philips Medical Systems (Cleveland) Inc · posted Nov 1, 2019 · Terminated
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
- Z-0214-2020 · Philips Medical Systems (Cleveland) Inc · posted Nov 1, 2019 · Terminated
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
- Z-0215-2020 · Philips Medical Systems (Cleveland) Inc · posted Nov 1, 2019 · Terminated
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
- Z-0216-2020 · Philips Medical Systems (Cleveland) Inc · posted Nov 1, 2019 · Terminated
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
- Z-0217-2020 · Philips Medical Systems (Cleveland) Inc · posted Nov 1, 2019 · Terminated
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
Class II recalls (16)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0574-2020 · Abbott Gmbh & Co. KG · posted Nov 26, 2019 · Terminated
Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a decrease in the control values and patient…
- Z-0548-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0549-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0550-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0551-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0552-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0553-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0554-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0555-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0556-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0557-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0558-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0559-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0560-2020 · Boston Scientific Corporation · posted Nov 25, 2019 · Terminated
Product manufactured between September 29, 2017 and September 12, 2019 have been labeled with a 36-month shelf life rather than the approved 18-month shelf life
- Z-0498-2020 · Philips North America, LLC · posted Nov 19, 2019 · Terminated
The rotary therapy selector switch may fail, resulting in unexpected device behavior. These behaviors include: The device may not turn on; The device may not perform the selected function; The device may deliver a shock with an energy level different from…
- Z-0364-2020 · Implant Direct Sybron Manufacturing, LLC · posted Nov 12, 2019 · Terminated
A QA inspection confirmed that a Healing Collar (Part number 8735-15) was packaged instead of the Implant ordered.
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0546-2020 · Spineart SA · posted Nov 22, 2019 · Terminated
Inner blister pack label is incorrect and does not match the correct external box label.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.