Elekta Unity, Image-Guided Radiation Therapy System recallZ-1980-2019 · Elekta, Inc. · initiated Jun 24, 2019

Class II Terminated All recalls posted in July 2019

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The QA software solution to perform the MR to MV alignment check, does not display the stored MR to MV offset values. The user is unable to independently inspect the values during their QA.

FDA root cause category: Software design

What the company is doing

The firm, Elekta, disseminated their "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" by email on 06/24/2019 to its customers. The notices identified the problem, provided recommended actions, and requested the notice be placed in a place accessible to all users. Customers were also instructed to complete and return the Acknowledgement Form. A software update will be provided once available. If you have any questions, contact the Post Market Surveillance Specialist by email: [email protected].

Recall details

Recall number
Z-1980-2019
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 24, 2019
Posted by the FDA
Jul 16, 2019
Recalling firm
Elekta, Inc.
Quantity
12 units
Product code
IYE
FDA submissions
K182076
Distribution
US Distribution to states of: TX & WI, and Internationally to: Canada, Denmark, Germany, Italy, Netherlands, Sweden, and the UK.

Product description

Elekta Unity, Image-Guided Radiation Therapy System Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.

Code information (models, lots, serial numbers)

UDI/GTIN: 5060081071321 Installed product: UNITY SYSTEM/13553-006/600003 UNITY SYSTEM/10817-011/600010 UNITY SYSTEM/12473-020/600020 UNITY SYSTEM/12274-017/600012 UNITY SYSTEM/10562-011/600016 UNITY SYSTEM/10719-T03001/600013 UNITY SYSTEM/30004070-001/600023 UNITY SYSTEM/10420-AVL-U/600008 UNITY SYSTEM/11014-UMCU-U/600007 UNITY SYSTEM/10157-006/600014 UNITY SYSTEM/11611-40/600011 UNITY SYSTEM/11489-21/600009

Other devices under the same FDA submission

1 device cleared under K182076 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.