Skip to content

BD Vacutainer PST Gel and Lithium HeparinN (LH) 56 Units… recallZ-2307-2019 · Becton Dickinson & Company · initiated Jun 12, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.

FDA root cause category: Device Design

What the company is doing

Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers.

Recall details

Recall number
Z-2307-2019
Status
Terminated
Initiated by the company
Jun 12, 2019
Recalling firm
Becton Dickinson & Company
Quantity
1,436,228,004 total
Product code
JKA
FDA submissions
K944566
Distribution
Worldwide Distribution, including US Nationwide

Product description

BD Vacutainer PST Gel and Lithium HeparinN (LH) 56 Units Blood Collection Tubes, Catalog Number 368056

Code information (models, lots, serial numbers)

UDI: (01)30382903680567

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

5 devices cleared under K944566 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 82963).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page