Skip to content

Maquet Cardiopulmonary (MCP)/Getinge HLS Set Advanced 7.0.… recallZ-2149-2019 · Maquet Cardiovascular Us Sales, Llc · initiated Jul 11, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Cardiovascular Us Sales, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The sets are configured with quick connectors that have been assembled in reverse on the arterial and venous lines, resulting in blue (venous) to red (arterial) and red (arterial) to blue (venous) connections. Due to this incorrect assembly, the sets cannot be primed from the reservoir prior to use, and the device may not function as intended.

FDA root cause category: Under Investigation by firm

What the company is doing

Urgent Medical Device Recall - Removal notification letters dated 7/11/19 were sent to customers.

Recall details

Recall number
Z-2149-2019
Status
Terminated
Initiated by the company
Jul 11, 2019
Recalling firm
Maquet Cardiovascular Us Sales, Llc
Quantity
54
Product code
DTZ
FDA submissions
K112360
Distribution
The products were distributed to the following US states: AR, CA, FL, GA, IA, IL, IN, KY, NC, NE, NM, NV, NY, OH, OR, TN, and WI.

Product description

Maquet Cardiopulmonary (MCP)/Getinge HLS Set Advanced 7.0. Model Number BEQ-HLS-7050 USA, Part Number 701052794

Code information (models, lots, serial numbers)

Lot 70131093; UDI 04037691741543

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K112360 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page