GammaPod Stereotactic Radiotherapy System Model A, REF… recallZ-2081-2019 · Xcision Medical Systems, LLC · initiated Jul 3, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Xcision Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The firm sent a Field Correction Notice letters dated 7/3/19 to customers. The letter identified the affected product, problem and actions to be taken. If you need any further information or support concerning this issue, please contact Xcsion Customer Support at 855-924-7466.

Recall details

Recall number
Z-2081-2019
Status
Terminated
Initiated by the company
Jul 3, 2019
Recalling firm
Xcision Medical Systems, LLC
Quantity
2
Product code
IWB
FDA submissions
K180571
Distribution
The products were distributed to the following US states: MD and TX.

Product description

GammaPod Stereotactic Radiotherapy System Model A, REF XMSGP030A00-0.02 Product Usage: GammaPod is a teletherapy device intended for use in the noninvasive stereotactic delivery of a radiation dose to a partial volume of the breast in conjunction with breast conserving treatment.

Code information (models, lots, serial numbers)

Serial Numbers GP0002, GP0004; UDI: (01) 00850002615005

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.