Skip to content

GammaPod™ Model ARecalls & reported problems

Xcision Medical Systems, LLC · Stereotactic Radiotherapy System · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Nov 30, 2023Open, Classified

    Wrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS generated from a previous CT scan to be selected

    Z-0750-2024 · Root cause: Software design · FDA record ↗

  2. Jul 3, 2019Terminated

    One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.

    Z-2081-2019 · Root cause: Nonconforming Material/Component · FDA record ↗

Other System, Radiation Therapy, Radionuclide devices

All System, Radiation Therapy, Radionuclide devices →