Why it was recalled

Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE onwards).

FDA root cause category: Under Investigation by firm

What the company is doing

Randox Laboratories Ltd. issued Urgent Medical Device Correction letter on 7/12/19 advising users of the problem, health risk and action to take: Replace all Quality Control value sheets with the revised sheets available on www.randox.com . Discuss the contents of this notice with your Medical Director. Complete and return the response form 12187-QA to [email protected] within five working days.

Recall details

Recall number
Z-2490-2019
Status
Terminated
Initiated by the company
Jul 12, 2019
Recalling firm
Randox Laboratories Ltd.
Quantity
307 kits
Product code
JIX
FDA submissions
K053153
Distribution
Nationwide Foreign: Antigua, Aruba, Bahamas, Granada, Tobago,

Product description

This is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. Randox calibration sera contain 42 analytes and are based on lyophilized human serum. The concentrations and activities are suitable for calibration of clinical chemistry assays both manually and on a wide range of automatic analyzers. Constituent concentrations are available at 2 levels.

Code information (models, lots, serial numbers)

Lot Numbers: 961UE onwards GTN: 05055273200966