Hemashield recallZ-2559-2019 · Getinge US Sales LLC · initiated Aug 22, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Getinge US Sales LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Based on an internal incident report, Getinge/Intervascular SAS has identified that the one (1) HEMASHIELD GOLD Knitted Microvel Double Velour Vascular Bifurcated Graft product with the diameter size of 16 x 9 mm is likely incorrectly packed inside of the carton which was labeled to contain a HEMASHIELD PLATINUM Woven Double Velour Vascular Graft product with the diameter size of 16 x 8 mm ((REF: M00202166168P0, Lot: 18K31, SN: 1190443415, size 16 x 8 mm).
FDA root cause category: Under Investigation by firm
What the company is doing
1. Please examine your inventory immediately to determine if you have the HEMASHIELD PLATINUM Woven Double Velour Vascular Graft product. 2. Please remove the affected product from areas of use. 3. Affected product should be returned to Getinge per the following process: Please complete the Urgent Medical Device Recall Removal Response Form to acknowledge that you have received this Urgent Medical Device Recall Removal letter. Please fax or email the completed form to Getinge office as instructed on the form. Please contact Getinge Customer Service at 18886278383 (press option 2, then option 5) Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone), to request a shipping instructions to return any affected product. Pack the product to be returned with the appropriate return documents. If you have any questions, please contact your Getinge representative or call the Getinge Customer Support at +1 8886278383 (press option 2, then option 5), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone)
Recall details
- Recall number
- Z-2559-2019
- Status
- Terminated
- Initiated by the company
- Aug 22, 2019
- Recalling firm
- Getinge US Sales LLC
- Quantity
- 1 unit
- Product code
- MAL
- FDA submissions
- K021213
- Distribution
- US state of NJ
Product description
The HEMASHIELD PLATINUM Woven Double Velour Vascular Graft is a woven double velour polyester graft impregnated with highly purified bovine collagen. The HEMASHIELD PLATINUM Woven Double Velour Vascular Graft is designed to reduce bleeding at implant and thereby eliminates the operative preclotting step, including cumbersome autoclave techniques. The collagen is designed to be gradually resorbed by the patient. The CONCENTRICRIMP pleat and GUIDELINE stripe are also featured. In addition to collagen, the graft also contains glycerol as a softening agent.
Code information (models, lots, serial numbers)
Lot Code: 18K31 Serial Number: 1190443415 Part Number: M00202166168P0 UDI Code: 00384401018452
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Hemashield M00202166168P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite1 recall3 reports
Product lines
Other devices under the same FDA submission
28 devices cleared under K021213 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Hemashield M00202175928P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite10 reports
- Hemashield M00202175208P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- Hemashield M00202175210P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite8 reports
- Hemashield M00202175430P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite6 reports
- Hemashield M00202175132P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite5 reports
- Hemashield M00202175410P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite5 reports
- Hemashield M00202175926P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite5 reports
- Hemashield M00202175428P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite4 reports
- Hemashield M00202175826P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite4 reports
- Hemashield M00202175830P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite4 reports
- Hemashield M00202175924P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite4 reports
- Hemashield M00202175126P0Intervascular Sas · Graft, Vascular, Synthetic/Biologic Composite3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page