Why it was recalled

Interference can occur between the adolescent attachment bolt and the adolescent nail, cause a delay in surgical procedures

FDA root cause category: Device Design

What the company is doing

OrthoPediatrics, Inc. notified Distributors by letter via email on 8/19/19 , stating reason for recall, health risk and action to take: Distributors have been instructed to review their inventory and the inventory at their clinical sites to recover the product. Questions contact: Kriss Anderson, VP of Quality Systems and Regulatory Affairs, at OrthoPediatrics at [email protected] or 574-267-0897.

Recall details

Recall number
Z-0125-2020
Status
Terminated
Initiated by the company
Aug 19, 2019
Recalling firm
OrthoPediatrics Corp
Quantity
211 units
Product code
HSB
FDA submissions
K172583
Distribution
Nationwide Foreign: Australia

Product description

OrthoPediatrics Adolescent attachment bolt used with the Pediatric Nailing Platform (PNP) targeting arm- intended as an orthopedic implant system (bolt is used to attach the targeting arm to the intramedullary fixation rod) Part number: 01-1503-0032

Code information (models, lots, serial numbers)

Lot/serial number: 192432-B 192279-B 189808-A 187375-A 177577-A 183280-A 175221-A 174784-A UDI: 00841132149836

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

9 devices cleared under K172583 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.