GUIDE WIRE 2.2MM X 28IN, Aim/Art Guide wire, Item No.… recallZ-0529-2020 · Zimmer Biomet, Inc. · initiated Aug 21, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.
FDA root cause category: Environmental control
What the company is doing
On September 11, 2019, the firm began notifying distributors and customers of the recall via an Urgent Medical Device Recall letter. The letter informed consignees of the product issue. Customers were asked to assist their Zimmer Biomet sales representative and quarantine all identified product. The sales representative will remove the product from your facility. If any products were further distributed, please forward the notification to responsible parties. If you have further questions or concerns, please call customer service at 574-371-3071 between 8:00 am and 5:00 pm ET, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule.
Recall details
- Recall number
- Z-0529-2020
- Status
- Terminated
- Initiated by the company
- Aug 21, 2019
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 2468
- Product code
- FZX
- Distribution
- US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.
Product description
GUIDE WIRE 2.2MM X 28IN, Aim/Art Guide wire, Item No. 809222028 - Product Usage: These instruments and delivery systems are used to facilitate implantation of orthopedic medical devices.
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Lots: 323740 323770 323870 323970 324050 957830 957880 323840 323890 323910 323930 957800 987260 987480 987720 485140 486100 485470 485850 069570 069980 070100 977580 977670 977690 977700 977710 977720 977730 977750 661460 661440 661450 161370 166860 166870 166870R 108410 108350 108380 108390 108400 108430 108450 472150 392630 392650 392660 392680 233870 233900 471850 471860 471870 472140 472160 036430 036390 036370 036360 036400 036450 036380 036410 036420 036440 036460 121498 212067 270142 270144 270146 270149 270151 270153 A3WAA7 A3WAB7 A3WAC7 A4WAA7 A4WAC7 A4WAD7 AD4AB7000 ADJAB7000 AP8AC7 AP8AD7000 B3TAH7 BA8AA7 BY3AB7 C25AA7 C25AB7 C4YAA7 CA3AA7 CD8AA7 CW3AA7 CW3AB7 CW3AC7 D38AC7 D3CAC7 D3CAF7 D4EAD7 D4EADA D5VAJ7 D66AD7 DF9AA7 DH6AA7 DH6AA7R DN5AJ7 DV3AA7 DW7AA7 DW7AAA E2PAD7 E2PAE7 E2PAG7 E2YAD7 E36AC7 E45AD7 E4EAB7 E55AD7 E55AE7 E5CAB7 E5CAC7 E5VAJ7 EG9AA7 EP2AC7 EP9AC7 ES5AA7 EY2AC7 EY9AA7 FB1AA7 FB1AB7 FB8AB7 FB8AC7 W35AD7000 WS1AJ7000 XE7AW7000 Y1KAB7000 Y98AA7000 YJ7AA7000 Z78AA7000 ZJ2AA7000 ZJ5AA7000 ZJ5AAA UDI: (01)00887868034664(17)240619(10)323970 (01)00887868034664(17)240619(10)324050 (01)00887868034664(17)240708(10)957800 (01)00887868034664(17)240716(10)987260 (01)00887868034664(17)240717(10)987480 (01)00887868034664(17)240717(10)987720 (01)00887868034664(17)240828(10)485140 (01)00887868034664(17)240828(10)486100 (01)00887868034664(17)240829(10)485470 (01)00887868034664(17)240902(10)485850 (01)00887868034664(17)241015(10)069570 (01)00887868034664(17)241016(10)069980 (01)00887868034664(17)241017(10)070100 (01)00887868034664(17)250121(10)977580 (01)00887868034664(17)250122(10)977670 (01)00887868034664(17)250122(10)977690 (01)00887868034664(17)250122(10)977700 (01)00887868034664(17)250122(10)977710 (01)00887868034664(17)250122(10)977720 (01)00887868034664(17)250122(10)977730 (01)00887868034664(17)250122(10)977750 (01)00887868034664(17)250408(10)661460 (01)00887868034664(17)250408(10)661440 (01)00887868034664(17)250408(10)661450 (01)00887868034664(17)250601(10)161370 (01)00887868034664(17)250630(10)166860 (01)00887868034664(17)250601(10)166870 (01)00887868034664(17)250601(10)166870R (01)00887868034664(17)250902(10)108410 (01)00887868034664(17)250902(10)108350 (01)00887868034664(17)250902(10)108380 (01)00887868034664(17)250902(10)108390 (01)00887868034664(17)250902(10)108400 (01)00887868034664(17)250903(10)108430 (01)00887868034664(17)250903(10)108450 (01)00887868034664(17)260610(10)472150 (01)00887868034664(17)260630(10)392630 (01)00887868034664(17)260111(10)392650 (01)00887868034664(17)260111(10)392660 (01)00887868034664(17)260111(10)392680 (01)00887868034664(17)260111(10)233870 (01)00887868034664(17)260111(10)233900 (01)00887868034664(17)260616(10)471850 (01)00887868034664(17)260111(10)471860 (01)00887868034664(17)260610(10)471870 (01)00887868034664(17)260616(10)472140 (01)00887868034664(17)260623(10)472160 (01)00887868034664(17)260614(10)036430 (01)00887868034664(17)260616(10)036390 (01)00887868034664(17)260623(10)036370 (01)00887868034664(17)260624(10)036360 (01)00887868034664(17)260624(10)036400 (01)00887868034664(17)260624(10)036450 (01)00887868034664(17)260628(10)036380 (01)00887868034664(17)260628(10)036410 (01)00887868034664(17)260628(10)036420 (01)00887868034664(17)260628(10)036440 (01)00887868034664(17)260628(10)036460
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- GUIDE WIRE 8092-22-028Biomet Orthopedics, LLC · Guide, surgical, instrument1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 83783).
- Z-0519-2020 · TerminatedCompress, Mini Compress; Item Nos. 178350 178351 178352 178353 178354 178355 178356 178357 178358 178359 178360 178361 178362 178363 178364 178365 178366 178367 178368 178369 17837…
- Z-0520-2020 · TerminatedCOMP RVRS Shoulder Baseplates, including Custom Products, Item numbers 115331S 115331 115330S 115330 110040610 110040620 010000589 CP561861 PM0001456 PM551047 PM551063 PM555137 PM5…
- Z-0523-2020 · TerminatedVarious Hip systems, Item Nos. 104207 104212 104215 11-162809 11-162810 11-162811 11-162812 11-162813 21-103202 21-103203 21-103204 21-103205 21-103206 21-103207 21-103208 21-10320…
- Z-0526-2020 · TerminatedVarious Custom Products, Item Nos. CP102940 CP102941 CP102942 CP102943 CP102944 CP102945 CP102946 CP102947 CP102948 CP112662 CP112663 CP112664 CP112674 CP112675 CP113771 CP113777 C…
- Z-0527-2020 · TerminatedCopeland HA Resurfacing Head, MB/HA EAS Head; Item Nos. 11-114641 11-114642 11-114643 11-114644 11-114644S 11-114645 11-114646 11-114647 11-114648 11-114661 11-114662 11-114663 11-…
- Z-0528-2020 · TerminatedHA Pins (DFS HA Cort Screw), Hoffman/Colles Pins (Colles CSER Sterile w/ Pin Kit) Item Nos: HAA60-09030 HAA60-09040 HAA60-10030 HAA60-10040 HAA60-11030 HAA60-11040 HAA60-11050 HAA…
- Z-0530-2020 · TerminatedBALL NOSE GUIDE WIRE 80CM, 100CM. Item Nos. 281001080 281001100 Product Usage: These instruments and delivery systems are used to facilitate implantation of orthopedic medical dev…
- Z-0531-2020 · Terminated2.0MM BALL NOSE GUIDE WIRE, Item Nos. 281017006 - Product Usage: These instruments and delivery systems are used to facilitate implantation of orthopedic medical devices.
- Z-0533-2020 · TerminatedBIOMOD PORCOAT HA GLENTRAY; Custom Shoulder Products Item Nos. PM0000097 PM555140 PM555147 PM555319 PM555418 113867 113865 113863 Product Usage: Total Shoulder Replacement
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page