BALL NOSE GUIDE WIRE recallZ-0531-2020 · Zimmer Biomet, Inc. · initiated Aug 21, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.

FDA root cause category: Environmental control

What the company is doing

On September 11, 2019, the firm began notifying distributors and customers of the recall via an Urgent Medical Device Recall letter. The letter informed consignees of the product issue. Customers were asked to assist their Zimmer Biomet sales representative and quarantine all identified product. The sales representative will remove the product from your facility. If any products were further distributed, please forward the notification to responsible parties. If you have further questions or concerns, please call customer service at 574-371-3071 between 8:00 am and 5:00 pm ET, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule.

Recall details

Recall number
Z-0531-2020
Status
Terminated
Initiated by the company
Aug 21, 2019
Recalling firm
Zimmer Biomet, Inc.
Quantity
2708
Product code
FZX
Distribution
US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.

Product description

2.0MM BALL NOSE GUIDE WIRE, Item Nos. 281017006 - Product Usage: These instruments and delivery systems are used to facilitate implantation of orthopedic medical devices.

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

LOTS:  206322 062980 062690 062720 062740 062760 062950 062960 062970 730390 730400 730380 158530 158510 158520 460220 859610 860250 661640 977840 977850 114600 167450 204150 204160 204180 204190 204200 204210 678430 491410 678420 233730 233770 392590 233710 233720 694090 039330 039340 039350 039360 039370 039380 039450 039420 039390 039400 039410 039430 039440 039460 039470 113340 994740 036320 036330 665740 994730 036290 204970 204990 205000 205010 205020 472130 164579 164582 164586 164587 206318 206319 206320 206321 260820 260821 270166 270167 AF6AE7000 AT9AG7 B5KAT7 B5LAA7 B78AE7 B78AF7 BD3AA7 BT9AD7 BX2AB7 CN8AG7 CN8AH7 CW3AT7 CW3AV7 CW3AV7R CW3AW7 CW3AW7R CW3AX7 E3SAB7 E3SAC7 EG9AF7 EG9AG7 EG9AH7 EG9AJ7 NXW351U NXW351V  UDI:  (01)00887868029448(17)220501(10)206322 (01)00887868029448(17)240819(10)062980 (01)00887868029448(17)240819(10)062690 (01)00887868029448(17)240819(10)062720 (01)00887868029448(17)240819(10)062740 (01)00887868029448(17)240819(10)062760 (01)00887868029448(17)240819(10)062950 (01)00887868029448(17)240819(10)062960 (01)00887868029448(17)240819(10)062970 (01)00887868029448(17)241013(10)730390 (01)00887868029448(17)241013(10)730400 (01)00887868029448(17)241013(10)730380 (01)00887868029448(17)241101(10)158530 (01)00887868029448(17)241103(10)158510 (01)00887868029448(17)241103(10)158520 (01)00887868029448(17)241201(10)460220 (01)00887868029448(17)250115(10)859610 (01)00887868029448(17)250116(10)860250 (01)00887868029448(17)250410(10)661640 (01)00887868029448(17)250429(10)977840 (01)00887868029448(17)250429(10)977850 (01)00887868029448(17)250601(10)114600 (01)00887868029448(17)250601(10)167450 (01)00887868029448(17)250921(10)204150 (01)00887868029448(17)250921(10)204160 (01)00887868029448(17)250921(10)204180 (01)00887868029448(17)250921(10)204190 (01)00887868029448(17)250922(10)204200 (01)00887868029448(17)250922(10)204210 (01)00887868029448(17)260614(10)678430 (01)00887868029448(17)260623(10)491410 (01)00887868029448(17)260108(10)678420 (01)00887868029448(17)260111(10)233730 (01)00887868029448(17)260111(10)233770 (01)00887868029448(17)260616(10)392590 (01)00887868029448(17)260111(10)233710 (01)00887868029448(17)260111(10)233720 (01)00887868029448(17)260403(10)694090 (01)00887868029448(17)260502(10)039330 (01)00887868029448(17)260502(10)039340 (01)00887868029448(17)260502(10)039350 (01)00887868029448(17)260502(10)039360 (01)00887868029448(17)260505(10)039370 (01)00887868029448(17)260505(10)039380 (01)00887868029448(17)260524(10)039450 (01)00887868029448(17)260525(10)039420 (01)00887868029448(17)260525(10)039390 (01)00887868029448(17)260525(10)039400 (01)00887868029448(17)260525(10)039410 (01)00887868029448(17)260525(10)039430 (01)00887868029448(17)260525(10)039440 (01)00887868029448(17)260525(10)039460 (01)00887868029448(17)260525(10)039470 (01)00887868029448(17)260610(10)113340 (01)00887868029448(17)260610(10)994740 (01)00887868029448(17)260623(10)036320 (01)00887868029448(17)260623(10)036330 (01)00887868029448(17)260623(10)665740 (01)00887868029448(17)260623(10)994730 (01)00887868029448(17)260701(10)036290 (01)00887868029448(17)260718(10)204970 (01)00887868029448(17)260718(10)204990 (01)00887868029448(17)260718(10)205000 (01)00887868029448(17)260718(10)205010 (01)00887868029448(17)260718(10)205020

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 83783).