The Infinity Acute Care System (IACS) Monitoring Solution… recallZ-0258-2020 · Draegar Medical Systems, Inc. · initiated Sep 13, 2019

Class II Open, Classified All recalls posted in October 2019

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draegar Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Cybersecurity vulnerabilities may cause device to reboot, lose alarm functionality, and/or lose communication with cockpit and/or the Infinity Network.

FDA root cause category: Software design

What the company is doing

On September 13, 2019, the firm distributed Urgent Medical Device Recall letters to customers. Customers were informed of the cybersecurity vulnerabilities, which can include Distributed Denial of Service (DDOS) (packet storm), Spoofing, and Tampering. To mitigate the cybersecurity concerns, Draeger will be releasing software version VG4.2 for both the Cockpit and the M540, which will correct these cybersecurity vulnerabilities. The software is expected to be released for distribution in December 2019. Upgrades of the IACS systems will commence in January 2020. While the firm is in the process of updating the software, customers are asked to limit access to the Infinity Network by following these security recommendations: 1. Physical security of the patient monitors is recommended and is the responsibility of the operating organization. 2. Physical security of the telecommunications closet is recommended and is the responsibility of the operating organization. 3. Draeger recommends that operating organizations restrict physical access to unused Ethernet ports on the IACS. 4. Draeger recommends that operating organizations restrict physical access to unused USB and serial ports on the IACS. 5. Draeger relies on the medical device isolation mechanism of the VLANs and the proper configuration, implementation, and use of the operating organization's security measures to prevent the introduction of malware onto the Infinity Network. Your local Draeger Service Representative will contact you to schedule an appointment to upgrade your system(s) software free of charge once the new software version is released for distribution. If you have any questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM - 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).

Recall details

Recall number
Z-0258-2020
Recall class
Class II
Status
Open, Classified
Initiated by the company
Sep 13, 2019
Posted by the FDA
Oct 31, 2019
Recalling firm
Draegar Medical Systems, Inc.
Quantity
5634
Product code
MHX
FDA submissions
K113798, K093788
Distribution
US Nationwide distribution.

Product description

The Infinity Acute Care System (IACS) Monitoring Solution with the Standalone Infinity M540 patient monitor. Model nos. MS25510, MS25520, MS26372.

Code information (models, lots, serial numbers)

Software version VG4.1.1/VG4.0.3 and lower

Other devices under the same FDA submission

5 devices cleared under K113798, K093788 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.