The Infinity Acute Care System (IACS) Monitoring Solution… recallZ-0258-2020 · Draegar Medical Systems, Inc. · initiated Sep 13, 2019
Class II Open, Classified All recalls posted in October 2019
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Draegar Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Cybersecurity vulnerabilities may cause device to reboot, lose alarm functionality, and/or lose communication with cockpit and/or the Infinity Network.
FDA root cause category: Software design
What the company is doing
On September 13, 2019, the firm distributed Urgent Medical Device Recall letters to customers. Customers were informed of the cybersecurity vulnerabilities, which can include Distributed Denial of Service (DDOS) (packet storm), Spoofing, and Tampering. To mitigate the cybersecurity concerns, Draeger will be releasing software version VG4.2 for both the Cockpit and the M540, which will correct these cybersecurity vulnerabilities. The software is expected to be released for distribution in December 2019. Upgrades of the IACS systems will commence in January 2020. While the firm is in the process of updating the software, customers are asked to limit access to the Infinity Network by following these security recommendations: 1. Physical security of the patient monitors is recommended and is the responsibility of the operating organization. 2. Physical security of the telecommunications closet is recommended and is the responsibility of the operating organization. 3. Draeger recommends that operating organizations restrict physical access to unused Ethernet ports on the IACS. 4. Draeger recommends that operating organizations restrict physical access to unused USB and serial ports on the IACS. 5. Draeger relies on the medical device isolation mechanism of the VLANs and the proper configuration, implementation, and use of the operating organization's security measures to prevent the introduction of malware onto the Infinity Network. Your local Draeger Service Representative will contact you to schedule an appointment to upgrade your system(s) software free of charge once the new software version is released for distribution. If you have any questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM - 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).
Recall details
- Recall number
- Z-0258-2020
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 13, 2019
- Posted by the FDA
- Oct 31, 2019
- Recalling firm
- Draegar Medical Systems, Inc.
- Quantity
- 5634
- Product code
- MHX
- FDA submissions
- K113798, K093788
- Distribution
- US Nationwide distribution.
Product description
The Infinity Acute Care System (IACS) Monitoring Solution with the Standalone Infinity M540 patient monitor. Model nos. MS25510, MS25520, MS26372.
Code information (models, lots, serial numbers)
Software version VG4.1.1/VG4.0.3 and lower
Other devices under the same FDA submission
5 devices cleared under K113798, K093788 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SHP UNT INFINITY M540 MS20401Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)2 recalls5 deaths62 reports
- IACS w/C700 MS25520Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)2 recalls1 death27 reports
- IACS w/C500 MS25510Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)2 recalls2 deaths25 reports
- Infinity M540 MS26372Draeger Medical Systems, Inc. · System,Network And Communication,Physiological Monitors2 recalls9 reports
- Infinity Mpod Quad Hemo MS20725Draeger Medical Systems, Inc. · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page