Atellica CH 930 Analyzer recallZ-0581-2020 · Siemens Healthcare Diagnostics, Inc. · initiated Sep 23, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for calibration error resulting in QC failures after the calibration run with QC and patient results being 5 times higher.
FDA root cause category: Software design
What the company is doing
Siemens issued an Urgent medical Device Correction, (UMDC ASW19-06.A.US) and Urgent Field Safety Notification (UFSN ASW19-06.A.OUS) to all affected Customers via Certified Mail (domestic) and e-mail (international) beginning on 2019-09-23. It included the following instructions to the user: Actions to be Taken by the Customer 1. Do not order Reagent Pack Calibration (C0 Adjust) while the Reagent Lot calibration is in progress. 2. Always run QC after calibration before running patient samples. If QC is ordered with calibration and it recovers out of range high, the calibration status will be 'Awaiting Acceptance'. Reject and re-order the calibration. - Please review this letter with your Medical Director. - Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. - If you have received any complaints of illness or adverse events associated with the product listed immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product.
Recall details
- Recall number
- Z-0581-2020
- Status
- Terminated
- Initiated by the company
- Sep 23, 2019
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 1432
- Product code
- JJE
- FDA submissions
- K151767
- Distribution
- Distributed to accounts located in AL AZ AR CA CO FL GA IL IN IA KS KY LA MA MI MN MS NE NV NH NJ NM NY NC OH OR PA SC TN TX UT WA WV DC PR. Foreign distribution to AR AU AT BS BD BE BR BG CA CL CO CW CZ DK EG FI FR DE GR HU IN IR IE IL IT JP KW LV MY MX NL NG NO CN PK PH PL PT QA KR RO RU SA SG SK ZA ES SE CH TW TH TR AE GB UY VN.
Product description
Atellica CH 930 Analyzer
Code information (models, lots, serial numbers)
SMN 11067000, Software V1.20.0 and Lower, UDI 00630414002163
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls679 reports
Other devices under the same FDA submission
5 devices cleared under K151767 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Atellica CH Diluent 11099300Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall0 reports
- Atellica A-LYTE™ Integrated Multisensor Na+ / K+ / Cl- 11099315Siemens Healthcare Diagnostics Inc. · ELECTRODE, ION SPECIFIC, POTASSIUM25 reports
- Atellica Direct Load 11068008Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE3 reports
- Atellica CH Cleaner 11099303Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 report
- Atellica CH AlbP 11097530Siemens Healthcare Diagnostics Inc. · BROMCRESOL PURPLE DYE-BINDING, ALBUMIN1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page