Atellica CH 930 Analyzer 11067000Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Atellica CH 930 Analyzer · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Siemens Atellica CH analyzer has 687 adverse event reports, with no deaths and 4 injuries. Most describe test results that were falsely low or high, such as microalbumin, urinary protein and carbon dioxide readings, some of which were retested on another analyzer. 5 recalls are on file, 2 of them open, covering software and valve problems that can affect results. Reports dropped, with 46 in the last year compared with 103 in the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 677
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 5
Summary
The FDA's MAUDE database contains 677 adverse event reports that list this exact device (DI 00630414002163), received between Jun 20, 2018 and Sep 3, 2026 — classified as 4 injuries, 673 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (256 reports).
5 FDA recall records name this device's identifier, 2 still open.
7 other recalls cover devices cleared under the same FDA submission (K151767); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Nov 18, 2021Class IIOpen, Classified
(1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay Correlation factor; (2) On Board Stability (OBS) Not Updating with Manual Change for Open Channel Assays may continue to use the reagent past its expiration date leading to potential erroneous patient sample results
Z-0449-2022 · Root cause: Software design · Recall details · FDA record ↗
- May 13, 2021Class IIOpen, Classified
Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject to high pressure or high wear and may impact the result accuracy of any assay.
Z-2026-2021 · Root cause: Component design/selection · Recall details · FDA record ↗
- Feb 27, 2020Class IITerminated
Editing an assay Test Definition and switching to another assay s Test Definition screen without saving the changes using the save button on the screen, may corrupt the settings of the assay s test definition, and Auto-Rerun or Auto-Dilution results may generate a result of Zero (0) for quantitative assays or Negative for qualitative assays if a well of the reagent pack is not calibrated.
Z-1797-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Oct 11, 2019Class IITerminated
On the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and Chloride (Cl), the system will translate the decimal point separators to commas. This will result in invalid concentration values for the fluids, therefore the system will utilize default nominal values instead of lot specific IMT fluid concentration values for calculation of the calibrations and results for the Na, K, and Cl methods.
Z-0862-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Sep 23, 2019Class IITerminated
Potential for calibration error resulting in QC failures after the calibration run with QC and patient results being 5 times higher.
Z-0581-2020 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Low Test Results256
- High Test Results201
- Incorrect, Inadequate or Imprecise Result or Readings129
- Application Program Problem: Parameter Calculation Error19
- Mechanical Problem16
- Imprecision12
- Adverse Event Without Identified Device or Use Problem7
- Computer Software Problem7
Patient problems
- No Clinical Signs, Symptoms or Conditions556
- No Known Impact Or Consequence To Patient118
- Chemical Exposure1
- Electric Shock1
- Laceration(s)1
- Needle Stick/Puncture1
Recent adverse event reports
The 25 most recent of 677 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2026Low Test Results
The customer reported that they obtained an erroneously depressed Calcium_2 (CA_2) result on a patient sample on an Atellica CH 930 Analyzer.The initial result was not reported to the physician(s). The sample was repeated on the same analyzer four times. The repeat results obtained were consistent with the patient¿s clinical picture and considered correct. A repeat result of 2.665 was reported to the physician(s) as correct result. There are no known reports of patient intervention or adverse health consequences due to the erroneously depressed CA_2 result.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50584370 ↗MalfunctionJun 23, 2026Low Test Results
THE CUSTOMER REPORTED THAT A FALSELY DEPRESSED MICROALBUMIN_2 (ALB_2) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPROCESSED ON THE ORIGINAL ANALYZER. THE REPEAT RESULT MATCHED THE PATIENT¿S CLINICAL ASSESSMENT. THE REPEAT RESULT WAS REPORTED AS CORRECT RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO A FALSELY DEPRESSED ALB_2 RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25595543 ↗MalfunctionMay 19, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER QUESTIONED THE ACCURACY OF THE INITIAL N LATEX FLC KAPPA RESULTS FOR A PATIENT SAMPLE TESTED ON THE ATELLICA CH 930 ANALYZER WITH SERIAL NUMBER (B)(6). THE CUSTOMER STATED THAT THE INITIAL RESULTS WERE NOT REPORTED TO THE PHYSICIANS AND THAT THE SAME SAMPLE AND ANALYZER WERE USED TO PERFORM REPEAT TESTING WITH A ×10 AUTO-DILUTION FACTOR. THE CUSTOMER CONSIDERED THE REPEAT RESULT OF 40.50 MG/L TO BE CORRECT AND REPORTED THE RESULT TO THE PHYSICIANS. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25217017 ↗MalfunctionMay 15, 2026Low Test Results
THE CUSTOMER REPORTED THAT A FALSELY DEPRESSED CALCIUM (CA) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN(S), WHO DID NOT QUESTION THE RESULT. THE SAMPLE WAS REPEATED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE REPEAT RESULT CORRELATED WITH PATIENT HISTORY/CLINICAL STATUS AND WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S).THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED CA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25180744 ↗MalfunctionMay 11, 2026Low Test Results
THE CUSTOMER REPORTED THAT THE ERRONEOUSLY DEPRESSED CARBON DIOXIDE, CONCENTRATED (CO2_C), ENZYMATIC CREATININE_3 (ECRE3) AND TOTAL BILIRUBIN_2 (TBIL_2) RESULTS WERE OBTAINED ON MULTIPLE PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. FOR THE CO2_C THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN(S). FOR ECRE3 AND TBIL_2 THE CUSTOMER DIDN'T PROVIDE THE INITIAL RESULTS. THEREFORE, THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN IS UNKNOWN. THE SAME SAMPLES WERE REPEATED ON THE ALTERNATE ATELLICA CH 930 ANALYZER¿S. THE REPEAT RESULTS CORRELATED WITH THE PATIENT¿S CLINICAL STATUS AND WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25133983 ↗MalfunctionApr 20, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED AN INACCURACY WITH TWO GENTAMICIN (GENT) RESULTS OBTAINED FOR ONE PATIENT ON THE ATELLICA CH 930 ANALYZER WITH SERIAL NUMBER (B)(6). THE INITIAL RESULT OF 0.64 MG/L WAS REPORTED TO THE PHYSICIANS. AFTER OBSERVING THAT QUALITY CONTROL (QC) WAS OUT OF RANGE AND REPLACING THE DILUTION PROBE, REPEAT TESTING WAS PERFORMED ON THE SAME ANALYZER, YIELDING A RESULT OF 1.85 MG/L. THE SAME PATIENT SAMPLE WAS THEN TESTED ON AN ALTERNATE ATELLICA CH 930 ANALYZER WITH ACCEPTABLE QUALITY CONTROL, AND THE RESULT OF 1.0 MG/L WAS CONSIDERED CORRECT. A CORRECTED REPORT WAS ISSUED AND COMMUNICATED TO THE PHYSICIANS. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24926755 ↗MalfunctionApr 17, 2026Low Test Results
THE CUSTOMER REPORTED FALSELY DEPRESSED ENZYMATIC CREATININE_3 (ECRE3) AND TOTAL BILIRUBIN_2 (TBIL_2) PATIENT RESULTS WERE OBTAINED ON AN ATELLICA CH 930 ANALYZER (S/N: (B)(6) V1.30.1). THE ERRONEOUS RESULTS WERE REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON ALTERNATE INSTRUMENTS. THE REPEAT RESULTS WERE REPORTED, AS THE CORRECT RESULTS, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24910119 ↗MalfunctionApr 9, 2026Low Test Results
ONE FALSELY DEPRESSED CALCIUM TEST RESULT WAS OBTAINED FOR ONE PATIENT SAMPLE FROM AN ATELLICA CH 930 ANALYZER. THE INITIAL RESULT WAS RELEASED TO A PHYSICIAN. THE SAME SAMPLE WAS TESTED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE RESULT FROM THE ALTERNATE ANALYZER WAS HIGHER, CONSIDERED CORRECT AND RELEASED TO A PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24834596 ↗MalfunctionApr 8, 2026High Test Results; Low Test Results
THE CUSTOMER REPORTED THAT THE ERRONEOUS CARBON DIOXIDE, CONCENTRATED (CO2_C) AND CREATININE_2 (CREA_2) RESULTS WERE OBTAINED ON MULTIPLE PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON THE ALTERNATE ATELLICA CH 930 ANALYZER¿S. THE REPEAT RESULTS CORRELATED WITH THE PATIENT¿S CLINICAL STATUS AND WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUS CO2_C AND CREA_2 RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24816620 ↗MalfunctionApr 7, 2026High Test Results
AN OUTSIDE OF THE UNITED STATES (OUS) CUSTOMER REPORTED THAT A FALSELY HIGH TEST RESULT FOR TOTAL BILIRUBIN (TBIL_2) WAS PRODUCED ON AN ATELLICA CH 930 ANALYZER. THE INITIAL TEST RESULT WAS RELEASED TO A PHYSICIAN. THE SAME SAMPLE WAS REPEATED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE TEST RESULT DATA FROM THE ALTERNATE ANALYZER WAS CONSIDERED CORRECT AND RELEASED TO A PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24802397 ↗MalfunctionApr 5, 2026Imprecise or Erratic Results
THE CUSTOMER REPORTED THAT FALSELY DEPRESSED CREATININE_2 (CREA_2) AND CREATINE KINASE (CK_L) RESULTS WERE OBTAINED ON A PATIENT SAMPLE, AND A FALSELY ELEVATED CARBON DIOXIDE, CONCENTRATED (CO2_C) RESULT WAS OBTAINED ON ANOTHER PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE REPEAT RESULTS MATCHED THE PATIENT¿S CLINICAL PICTURE. THE REPEAT RESULTS WERE DETERMINED TO BE CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUS CREA_2, CK_L, AND CO2_C RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24786638 ↗MalfunctionMar 30, 2026Low Test Results
THE CUSTOMER REPORTED THAT THEY OBTAINED ERRONEOUSLY DEPRESSED MICROALBUMIN_2 (ALB_2) RESULT ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULT WAS REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE REPEAT RESULT RECOVERED LOWER THAN THE INITIAL RESULT, CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED ALB_2 RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24711187 ↗MalfunctionMar 20, 2026High Test Results
THE CUSTOMER REPORTED THAT A FALSELY ELEVATED INORGANIC PHOSPHORUS (IP) RESULT FOR ONE PATIENT SAMPLE WAS OBTAINED ON THE ATELLICA CH 930 ANALYZER AND REPORTED TO PHYSICIANS. THE CUSTOMER USED THE SAME SAMPLE TO PERFORM REPEAT TESTING ON AN ALTERNATE ATELLICA CH 930 ANALYZER, WITH QUALITY CONTROL (QC) RESULTS THAT WERE ACCEPTABLE, AND NOTED THAT THE REPEAT RESULT WAS CORRECT. THE CUSTOMER ISSUED A CORRECTED REPORT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24649048 ↗MalfunctionMar 17, 2026Low Test Results
THE CUSTOMER REPORTED THAT THEY OBTAINED ERRONEOUSLY DEPRESSED CARBON DIOXIDE, CONCENTRATED (CO2_C) RESULTS ON THREE PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULTS WERE REPORTED TO THE PHYSICIAN(S) AND WERE QUESTIONED. THE SAMPLES WERE REPEATED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE REPEAT RESULTS RECOVERED HIGHER THAN THE INITIAL RESULT, CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED CO2_C RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24616831 ↗MalfunctionMar 12, 2026Sparking
A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) REPORTED THAT WHILE REMOVING THE QUICK-CONNECT TUBING ATTACHED TO THE LAMP/LIGHT-EMITTING DIODE (LED) ASSEMBLY, LAMP COOLANT SPILLED INTO THE DRIP TRAY IN THE CENTER REAR CABINET, OVERFLOWED ONTO THE 48-VOLT POWER SUPPLY, AND CAUSED A BURNING SMELL, A SPARK FROM THE POWER SUPPLY, AND SHUTDOWN OF AN ATELLICA CH 930 ANALYZER AT THE CUSTOMER¿S SITE. THERE WAS NO EXPOSURE TO BIOHAZARDOUS MATERIAL. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) AT THE TIME OF THE EVENT. THERE WAS NO SMOKE, FIRE, OR VISIBLE FLAME DUE TO THIS EVENT. THERE WAS NO INJURY TO THE OPERATOR DUE TO THIS EVENT. THERE ARE NO KNOWN REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24578420 ↗MalfunctionMar 11, 2026Low Test Results
THE CUSTOMER REPORTED THAT ERRONEOUSLY DEPRESSED CALCIUM (CA) RESULTS WERE OBTAINED ON THREE (3) PATIENT SAMPLES ON THE ATELLICA CH 930 ANALYZER. THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN(S) BUT NOT QUESTIONED. THE SAME SAMPLES WERE REPEATED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE REPEAT RESULTS WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED CA RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24566851 ↗MalfunctionMar 10, 2026High Test Results; Low Test Results
THE CUSTOMER REPORTED THAT ERRONEOUS CARBON DIOXIDE (CO2) AND CREATININE (CREA_2) RESULTS WERE OBTAINED ON TWO PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON DIFFERENT ALTERNATE ATELLICA CH 930 ANALYZERS. THE REPEAT RESULTS OBTAINED WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24557780 ↗MalfunctionMar 6, 2026High Test Results
THE CUSTOMER REPORTED THAT THEY OBTAINED ERRONEOUSLY ELEVATED ENZYMATIC HEMOGLOBIN A1C (A1C_E) RESULTS ON MULTIPLE PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULTS WERE REPORTED TO THE PHYSICIAN(S) AND WERE QUESTIONED. THE SAMPLES WERE REPEATED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE REPEAT RESULTS RECOVERED LOWER AND WERE CONSISTENT WITH THE PATIENT¿S CLINICAL HISTORY. THE CORRECTED REPORTS WERE ISSUED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY ELEVATED A1C_E RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24527901 ↗MalfunctionMar 5, 2026High Test Results; Low Test Results
THE CUSTOMER REPORTED THAT ERRONEOUS CARBON DIOXIDE, CONCENTRATED (CO2_C) AND CREATININE_2 (CREA_2) RESULTS WERE OBTAINED ON MULTIPLE PATIENT SAMPLES ON THE ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULTS WERE REPORTED TO THE PHYSICIAN(S) BUT NOT QUESTIONED. THE SAMPLES WERE REPEATED ON ALTERNATE ATELLICA CH 930 ANALYZERS. THE REPEAT RESULTS WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUS CO2_C AND CREA_2 RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24516761 ↗MalfunctionMar 2, 2026High Test Results
THE CUSTOMER REPORTED THAT A FALSELY ELEVATED CONCENTRATED CARBON DIOXIDE (CO2_C) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULT. THE SAMPLE WAS REPEATED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE REPEAT RESULT RECOVERED LOWER THAN THE INITIAL RESULT AND WAS CONSIDERED CORRECT. THE REPEAT RESULT WAS REPORTED AS A CORRECT RESULT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED CO2_C RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24492624 ↗MalfunctionMar 2, 2026Low Test Results
AN OUTSIDE OF THE UNITED STATES (OUS) CUSTOMER REPORTED A FALSELY LOW CREATININE (CREA_2) TEST RESULT WAS OBTAINED FOR ONE PATIENT SAMPLE FROM AN ATELLICA CH 930 ANALYZER. THE INITIAL RESULT WAS RELEASED TO A PHYSICIAN. THE SAME SAMPLE WAS REPEATED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. THE REPEAT RESULT WAS HIGHER AND RELEASED TO A PHYSICIAN AS THE CORRECT RESULT. THE ANALYZER POSTED A "PHOTOMETRIC LAMP INTENSITY OUT OF RANGE" ERROR WHILE THE TEST WAS PROCESSING. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE PHOTOMETRIC LAMP INTENSITY OUT OF RANGE ERROR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24489769 ↗MalfunctionMar 2, 2026High Test Results
THE CUSTOMER REPORTED AN ERRONEOUSLY ELEVATED CARBON DIOXIDE, CONCENTRATED (CO2_C) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULT WAS REPORTED TO THE PHYSICIAN(S) AND WAS QUESTIONED. THE SAME SAMPLE WAS REPEATED ON AN ALTERNATE ATELLICA CH ANALYZER. THE REPEAT RESULT WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY ELEVATED CARBON DIOXIDE, CONCENTRATED (CO2_C) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24480734 ↗MalfunctionFeb 18, 2026Low Test Results
THE CUSTOMER REPORTED THAT A FALSELY DEPRESSED SODIUM (NA) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE INITIAL RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED ON AN INITIAL ANALYZER. THE REPEAT RESULT RECOVERED HIGHER THAN THE INITIAL RESULT AND WAS CORRELATE WITH RESULT ON ALTERNATE ATELLICA CH 930 ANALYZER. THE REPEAT RESULT WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED NA RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24374849 ↗MalfunctionFeb 13, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED DISCORDANT CREATININE (CREA_2) RESULTS WERE OBTAINED ON MULTIPLE PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULTS WERE REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULT. THE SAMPLES WERE NOT REPEATED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24338326 ↗MalfunctionFeb 13, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED DISCORDANT CREATININE (CREA_2) RESULTS WERE OBTAINED ON MULTIPLE PATIENT SAMPLES ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUS RESULTS WERE REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULT. THE SAMPLES WERE NOT REPEATED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24338316 ↗
FDA clearance and classification
Cleared through 510(k) K151767 on Dec 11, 2015 (decision: Substantially Equivalent). Submitted by Siemens Healthcare Diagnostics, Inc. as “Trinidad CH system, TD-LYTE Integrated Multisensor (Na, K, CI), TD-LYTE IMT Standard A, TD-LYTE IMT Standard B + Salt Bridge, Trinidad CH Albumin BCP reagent (Alb_P), Trinidad CH Albumin BCP Calibrator”. The FDA review took 164 days. FDA record ↗
The FDA regulates product code JJE (analyzer, chemistry (photometric, discrete), for clinical use) under 21 CFR 862.2160 as a Class I device. 21 CFR 862.2160 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K151767. One submission often covers many models, so check the product and lot details in each record.
- Z-2027-2025 · May 22, 2025 · Open, ClassifiedProduct: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;
- Z-1003-2024 · Dec 15, 2023 · Open, ClassifiedTotal immunoglobulin M (IgM total) IVD, kit, nephelometry/turbidimetry
- Z-1870-2021 · May 4, 2021 · Open, ClassifiedAtellica CH 930 Analyzer with Photometer Lamp (SMN 11075676) Siemens Material Number: 11067000
- Z-0861-2021 · Dec 2, 2020 · TerminatedAtellica CH 930 Analyzer - In vitro diagnostic testing of clinical specimens. Product Siemens Material Number (SMN): 11067000 - Product Usage: intended for the qualitative and quan
- Z-0148-2020 · Jul 30, 2019 · TerminatedSiemens Atellica CH 930 Analyzer. Siemens Material Number (SMN): 11067000 Product Usage: automated, clinical chemistry analyzer to perform in vitro diagnostic tests on clinical
- Z-0782-2019 · Sep 26, 2018 · TerminatedAtellica CH 930 Analyzer. In vitro diagnostic testing of clinical specimens Siemens Material Number (SMN): 1 067000 - Product Usage: The Atellica Solution is a multi-component s
- Z-0194-2019 · Jun 29, 2018 · TerminatedAtellica CH 930 Analyzer (SMN 11067000) . The Atellica" Solution is a multi-component system for in vitro diagnostic testing of clinical specimens.
Common questions about the Atellica CH 930 Analyzer 11067000
Has the Atellica CH 930 Analyzer 11067000 been recalled?
Yes. 5 FDA recall records name this device's identifier: 2 still open and 3 closed (terminated) by the FDA. The most recent, Z-0449-2022 (Class II), began on Nov 18, 2021: (1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Atellica CH 930 Analyzer 11067000?
The FDA's MAUDE database lists 677 adverse event reports for this device, received between Jun 20, 2018 and Sep 3, 2026. In those reports, the most common specific device problems are “Low Test Results” (256), “High Test Results” (201) and “Incorrect, Inadequate or Imprecise Result or Readings” (129); the most common patient problems are “Chemical Exposure” (1), “Electric Shock” (1) and “Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Atellica CH 930 Analyzer 11067000?
None of the reports are classified as deaths, 4 as injuries and 673 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Atellica CH 930 Analyzer 11067000 FDA approved?
It was cleared through 510(k) K151767 on Dec 11, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Atellica CH 930 Analyzer 11067000, and is it still sold?
The FDA's device database lists Siemens Healthcare Diagnostics Inc. as the company responsible for it (brand Atellica CH 930 Analyzer, model 11067000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE devices
- Alinity 03R6701Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death1,184 reports
- Alinity 03R6501Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall903 reports
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
- ARCHITECT 3L77-01Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE2 recalls1 death692 reports
- ARCHITECT 3M74-02Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall650 reports
- ARCHITECT 2P24-40Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death645 reports
- cobas e 411 Immunoassay Analyzer 04775201001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE680 reports
- VITROS 6802413Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE12 recalls68 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page