Skip to content

Acidified Concentrate Distribution and Storage System (ACDS… recallZ-0821-2020 · Isopure Corp · initiated Apr 19, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Isopure Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The only affected customer was contacted via telephone on 4/19/19. An URGENT PRODUCT FIELD CORRECTION NOTICE dated 5/22/19 was distributed to affected clinics.

Recall details

Recall number
Z-0821-2020
Status
Terminated
Initiated by the company
Apr 19, 2019
Posted by the FDA
Jan 19, 2020
Recalling firm
Isopure Corp
Quantity
15
Product code
KPO
FDA submissions
K120005
Distribution
The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.

Product description

Acidified Concentrate Distribution and Storage System (ACDS System) with a Finish Thompson Centrifugal Pump Model #110192-3

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

ACDS UDI: (01)00852456007092; ACDS Serial Number AC-837874 with Pump SN 178745C18; ACDS Serial Number AC-834638 with Pump SN 180212D18; ACDS Serial Number AC-837654 with Pump SN 178324B18; ACDS Serial Number AC-837864 with Pump SN 178743C18; ACDS Serial Number AC-837664 with Pump SN 178322B18; ACDS Serial Number AC-838050 with Pump Serial Number 178752C18; ACDS Serial Number AC-838030 with Pump SN 178746C18; ACDS Serial Number AC-838020 with Pump SN 178748C18; ACDS Serial Number AC-838173 with Pump SN 178744C18; ACDS Serial Number AC-838143 with Pump SN 178325B18; ACDS Serial Number AC-838260 with Pump SN 180211D18; ACDS Serial Number AC-838740 with Pump SN 180209D18; ACDS Serial Number AC-838743 with Pump SN 180207D18; ACDS Serial Number AC-838749 with Pump SN 180215D18; ACDS Serial Number AC-838746 with Pump SN 180206D18

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 83990).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page