Medical device recalls: January 2020
The FDA posted 136 medical device recall records in January 2020, from 26 recall events by 22 companies. 9 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- GE Healthcare, LLC · 4 events, 14 records, 9 Class I
- Siemens Healthcare Diagnostics, Inc. · 2 events, 5 records
- Arrow International Inc · 1 event, 49 records
- Cardinal Health · 1 event, 39 records
- Flexicare Medical Ltd. · 1 event, 4 records
- Isopure Corp · 1 event, 3 records
- Siemens Healthcare Diagnostics, Inc · 1 event, 3 records
- Stryker Instruments Div. of Stryker Corporation · 1 event, 3 records
FDA root cause
- Other · 50
- Vendor change control · 39
- Under Investigation by firm · 12
- Process control · 12
- Nonconforming Material/Component · 8
- Software design · 6
- Device Design · 3
- Use error · 1
Class I recalls (9)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Carestation 620/650/650c A1Class IZ-0813-2020 · GE Healthcare, LLC · posted Jan 23, 2020 · Terminated
GE Healthcare has become aware that there is a potential for a loose cable connection inside specific manufactured anesthesia devices. This would cause a loss of mechanical ventilation and the system will provide high priority audio and visual alarms. Loss…
- Z-0814-2020 · GE Healthcare, LLC · posted Jan 23, 2020 · Terminated
GE Healthcare has become aware that there is a potential for a loose cable connection inside specific manufactured anesthesia devices. This would cause a loss of mechanical ventilation and the system will provide high priority audio and visual alarms. Loss…
- Z-0754-2020 · GE Healthcare, LLC · posted Jan 10, 2020 · Terminated
Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
- Z-0755-2020 · GE Healthcare, LLC · posted Jan 10, 2020 · Terminated
Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
- Z-0756-2020 · GE Healthcare, LLC · posted Jan 10, 2020 · Terminated
Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
- Z-0757-2020 · GE Healthcare, LLC · posted Jan 10, 2020 · Terminated
Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
- Z-0758-2020 · GE Healthcare, LLC · posted Jan 10, 2020 · Terminated
Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
- Z-0759-2020 · GE Healthcare, LLC · posted Jan 10, 2020 · Terminated
Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
- Airway Gas Option N-CAiO ↗Class IZ-0760-2020 · GE Healthcare, LLC · posted Jan 10, 2020 · Terminated
Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
Class II recalls (127)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft. ↗Class IIZ-1355-2020 · Cook Inc. · posted Jan 31, 2020 · Terminated
The firm has identified factors which contributed to failures relating to thrombus formation and/or lumen occlusion reports, and is highlighting to customers key points of the IFU pertaining to the prevention of these factors. In addition, the firm has…
- Z-0924-2020 · Jjgc Industria E Comercio De Materials Dentarios Sa · posted Jan 30, 2020 · Terminated
The laser engraving of the product incorrectly identifies the 4.3 Tapered Guided Surgery Drill as the "4.3+ Tapered Guided Surgery Drill".
- Z-1103-2020 · Fridababy LLC · posted Jan 30, 2020 · Open, Classified
FridaBaby has been made aware through customer reports that the product is being used outside of the labeled directions. Specifically, repeated chewing or biting on the SmileFrida, the ToothHugger has caused deterioration in the outer layer of the brush…
- Z-1363-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jan 30, 2020 · Terminated
Firm has received customer complaints regarding failed calibrations and increased imprecision of quality control and patient samples when using this product.
- Z-1347-2020 · Zimmer Biomet, Inc. · posted Jan 29, 2020 · Terminated
Potential increase in fracture rate was detected during internal testing, in addition to potential discoloration inside the inner pouch of the packaging.
- Z-0874-2020 · GE Healthcare, LLC · posted Jan 28, 2020 · Terminated
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date…
- Z-0875-2020 · GE Healthcare, LLC · posted Jan 28, 2020 · Terminated
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date…
- Z-0876-2020 · GE Healthcare, LLC · posted Jan 28, 2020 · Terminated
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date…
- Z-0877-2020 · GE Healthcare, LLC · posted Jan 28, 2020 · Terminated
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date…
- SafeAir Telescopic Smoke ↗Class IIZ-1097-2020 · Stryker Instruments Div. of Stryker Corporation · posted Jan 28, 2020 · Terminated
Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user.
- Telescopic Smoke Evac Pencil, ↗Class IIZ-1098-2020 · Stryker Instruments Div. of Stryker Corporation · posted Jan 28, 2020 · Terminated
Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user.
- Telescopic Uncoated (Push ↗Class IIZ-1099-2020 · Stryker Instruments Div. of Stryker Corporation · posted Jan 28, 2020 · Terminated
Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user.
- Zimmer Sterile Disposable ↗Class IIZ-1356-2020 · Zimmer Surgical Inc · posted Jan 28, 2020 · Terminated
When inflating/deflating a single bladder, both bladders could potentially inflate/deflate due to a leak between the bladders.
- Z-2045-2020 · Boston Scientific Corporation · posted Jan 28, 2020 · Terminated
A labeling discrepancy incorrectly labeled AMS 700 100mL Conceal Low Profile Reservoir with InhibiZone as 100mL Spherical Reservoir with InhibiZone.
- Z-0866-2020 · Raysearch Laboratories AB · posted Jan 27, 2020 · Terminated
Three issues found: i) The Map ROI options in the ROI list in the Structure Definition module may generate unintended ROI geometries ii) Elekta guard leaf behavior. There is an interoperability issue with Elekta regarding setting of guard leaves. iii)SSD,…
- Z-0867-2020 · Johnson & Johnson Surgical Vision Inc · posted Jan 27, 2020 · Terminated
Ophthalmic viscosurgical device may be difficult to remove from the eye, leading to increased postoperative intraocular pressure requiring additional intervention. Also, reports of clogging of phacoemulsion equipment tubing, which may lead to delay in the…
- Z-1227-2020 · CooperVision Inc. · posted Jan 27, 2020 · Terminated
Incorrect power of lenses than the expected
- Z-1228-2020 · CooperVision Inc. · posted Jan 27, 2020 · Terminated
Incorrect power of lenses than the expected
- Z-1269-2020 · Teleflex Medical Inc · posted Jan 27, 2020 · Terminated
There is a potential for the device packaging to become compromised; thus, the sterility of the device cannot be guaranteed.
- Z-1270-2020 · Teleflex Medical Inc · posted Jan 27, 2020 · Terminated
There is a potential for the device packaging to become compromised; thus, the sterility of the device cannot be guaranteed.
- Z-0612-2020 · Enztec Limited · posted Jan 24, 2020 · Terminated
The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response…
- Z-1002-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1003-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1004-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1005-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1006-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1007-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1008-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1009-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1010-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1011-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1012-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1013-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1014-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1015-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1016-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1017-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1018-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1019-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1020-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1021-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1022-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1023-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1024-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1025-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1026-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1027-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1028-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1029-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1030-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1031-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1032-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1033-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1034-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1035-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1036-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1037-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1038-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1039-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1040-2020 · Cardinal Health · posted Jan 24, 2020 · Terminated
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1170-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1171-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1172-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1173-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1174-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1175-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1176-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1177-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1178-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1179-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1180-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1181-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1182-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1183-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1184-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1185-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1186-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1187-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1188-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1189-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1190-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1191-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1192-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1193-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1194-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1195-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- FlexTip Plus Epidural Catheterization Set for Pediatric Lumbar Placement, Product Code EC-05520-P ↗Class IIZ-1196-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1197-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1198-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1199-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1200-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1201-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1202-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1203-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1204-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1205-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1206-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1207-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1208-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1209-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1210-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1211-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1212-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1213-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1214-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1215-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1216-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1217-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1218-2020 · Arrow International Inc · posted Jan 24, 2020 · Terminated
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
- Z-1230-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jan 24, 2020 · Terminated
There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.
- Z-1231-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jan 24, 2020 · Terminated
There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.
- Z-1232-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jan 24, 2020 · Terminated
There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.
- Z-1233-2020 · Siemens Healthcare Diagnostics, Inc. · posted Jan 24, 2020 · Terminated
There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.
- Z-0829-2020 · Siemens Healthcare Diagnostics, Inc · posted Jan 21, 2020 · Terminated
Homocysteine Assay May Cause Elevated Results in the Folate Assay
- Z-0830-2020 · Siemens Healthcare Diagnostics, Inc · posted Jan 21, 2020 · Terminated
Homocysteine Assay May Cause Elevated Results in the Folate Assay
- Siemens ADVIA Centaur Folate (500 Test Kit Reference)- for IVD of folate in serum or red blood cellClass IIZ-0831-2020 · Siemens Healthcare Diagnostics, Inc · posted Jan 21, 2020 · Terminated
Homocysteine Assay May Cause Elevated Results in the Folate Assay
- Z-0821-2020 · Isopure Corp · posted Jan 19, 2020 · Terminated
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion…
- Z-0822-2020 · Isopure Corp · posted Jan 19, 2020 · Terminated
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion…
- Z-0823-2020 · Isopure Corp · posted Jan 19, 2020 · Terminated
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion…
- Z-0819-2020 · GE Healthcare, LLC · posted Jan 17, 2020 · Terminated
Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
- Z-0801-2020 · Flexicare Medical Ltd. · posted Jan 13, 2020 · Terminated
The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could…
- Z-0802-2020 · Flexicare Medical Ltd. · posted Jan 13, 2020 · Terminated
The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could…
- Z-0803-2020 · Flexicare Medical Ltd. · posted Jan 13, 2020 · Terminated
The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could…
- Z-0804-2020 · Flexicare Medical Ltd. · posted Jan 13, 2020 · Terminated
The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could…
- Z-0795-2020 · Riverpoint Medical, LLC · posted Jan 10, 2020 · Terminated
The integrity of the sterile grid packaging is potentially compromised.
- Z-0782-2020 · Siemens Medical Solutions USA, Inc · posted Jan 7, 2020 · Terminated
During a routine inspection of a system, a crack on the C-arm holder was discovered. In the event the C-arm holder has a crack, the stability of the C-arm may be compromised and could lead to deformations of the C-arm. In very unlikely cases, an extreme…
- Z-0779-2020 · Medical Action Industries, Inc. 306 · posted Jan 6, 2020 · Terminated
The kits contain a bottle of saline solution which may exceed the USP specified pH requirements of 4.5-7.0.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.