Skip to content

ARCHITECT HAVAB-G Calibrator, REF 6L27-01 recallZ-0574-2020 · Abbott Gmbh & Co. KG · initiated Nov 7, 2019

Class II Terminated All recalls posted in November 2019

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Gmbh & Co. KG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a decrease in the control values and patient results. Patient values generated using the impacted calibrator lot 08219BE00 may have changed interpretation from reactive to falsely non reactive. Results impacted may present as a (falsely) non reactive result that fall in the range between 0.50 to 0.99 S/CO.

FDA root cause category: Material/Component Contamination

What the company is doing

The firm initiated the recall by letter on 11/07/2019. The letter explained the issue and requested that the consignee discontinue use of the product and destroy it. The firm is providing replacement material.

Recall details

Recall number
Z-0574-2020
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 7, 2019
Posted by the FDA
Nov 26, 2019
Recalling firm
Abbott Gmbh & Co. KG
Quantity
83 kits
Product code
LOL
FDA submissions
K113704
Distribution
US in the states of: AZ, CA, CO, CT, FL, KY, LA, MA, MD, MI, MN, MO, NJ, NY, OH, OR, RI, SC, TN, TX, WA, WV

Product description

ARCHITECT HAVAB-G Calibrator, REF 6L27-01

Code information (models, lots, serial numbers)

lot number 08219BE00

Other devices under the same FDA submission

2 devices cleared under K113704 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page