Skip to content

Radiofrequency Grounding Pad, REF RF-DGP-L recallZ-0664-2020 · Abbott Medical · initiated Oct 14, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected lots were manufactured with the protective release liner in an incorrect orientation. When attempting to remove this disposable release liner, a clear film may be left on the grounding pad which may interfere with patient skin contact and lead to uneven heating.

FDA root cause category: Process control

What the company is doing

The firm initiated the recall by letter hand delivered by an Abbott field representative. The letter requested that consignee cease use of the affected lot and return the product with the aid of a field representative.

Recall details

Recall number
Z-0664-2020
Status
Terminated
Initiated by the company
Oct 14, 2019
Recalling firm
Abbott Medical
Quantity
8940 devices
Product code
GXD
FDA submissions
K111576
Distribution
nationwide

Product description

Radiofrequency Grounding Pad, REF RF-DGP-L

Code information (models, lots, serial numbers)

UDI: 05415067028914 Batch Numbers: 810319004, 810319005, and 810319021

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page