NA RF-DGP-LRecalls & reported problems
St. Jude Medical, Inc. · Grounding Pad · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This St. Jude Medical electrosurgical grounding pad has 23 adverse event reports, mostly describing the pad's protective film being assembled the wrong way so a clear layer could remain on the pad and interfere with skin contact. Reports mention 2 injuries, including burns, and no deaths. The device was recalled in October 2019 for this manufacturing issue; that recall is now closed.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 23
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 05415067028914), received between Jun 25, 2019 and Oct 15, 2019 — classified as 2 injuries, 21 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (21 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Oct 14, 2019Terminated
Affected lots were manufactured with the protective release liner in an incorrect orientation. When attempting to remove this disposable release liner, a clear film may be left on the grounding pad which may interfere with patient skin contact and lead to uneven heating.
Z-0664-2020 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Deformation21
- Material Separation2
Patient problems
- No Patient Involvement18
- Burn(s)2
- No Consequences Or Impact To Patient2
- Superficial (First Degree) Burn1
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 15, 2019Material Deformation; Material Deformation
THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9191967 ↗MalfunctionOct 15, 2019Material Deformation; Material Deformation
THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9191955 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950659 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950646 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950644 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950643 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950619 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950612 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950588 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950582 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950570 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950550 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950543 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950533 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950508 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153 THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950501 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950488 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950486 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00132, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950465 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00131, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER.
Patient problems: No Patient Involvement
MAUDE report 8950440 ↗MalfunctionAug 30, 2019Material Deformation; Material Deformation
RELATED MANUFACTURING REF: 2182269-2019-00132, 2182269-2019-00134, 2182269-2019-00135, 2182269-2019-00136, 2182269-2019-00137, 2182269-2019-00138, 2182269-2019-00140, 2182269-2019-00141, 2182269-2019-00142, 2182269-2019-00143, 2182269-2019-00144, 2182269-2019-00145, 2182269-2019-00146, 2182269-2019-00147, 2182269-2019-00148, 2182269-2019-00149, 2182269-2019-00150, 2182269-2019-00153. THIS REPORT IS TO ADVISE OF AN EVENT DURING ANALYSIS CONFIRMING THE CLEAR FILM OF THE GROUNDING PAD WAS ASSEMBLED ON THE INSIDE LAYER OF THE BLUE LINER WHICH MAY HAVE RESULTED IN THE REPORTED FIRST DEGREE BURN THAT OCCURRED DURING A LEFT CERVICAL ABLATION PROCEDURE.
Patient problems: Superficial (First Degree) Burn
MAUDE report 8950430 ↗InjuryJun 25, 2019Material Separation; Material Separation
FOLLOWING A 4-LEVEL UNILATERAL LUMBAR ABLATION PROCEDURE, A PATIENT BURN WAS NOTED AFTER REMOVING THE GROUNDING PAD. THE GROUNDING PAD HAD LEFT A CLEAR FILM ON THE PATIENT FOLLOWING REMOVAL AND A BLISTER WAS NOTED ON THE SKIN UNDER THE LEFT FLANK. THE PATIENTS SKIN WAS NOT PREPPED BEFORE PLACING THE GROUNDING PAD. NO TREATMENT WAS NEEDED TO TREAT THE BURN AND THE PATIENT IS IN STABLE CONDITION.
Patient problems: Burn(s)
MAUDE report 8730591 ↗InjuryJun 25, 2019Material Separation; Material Separation
FOLLOWING A 4-LEVEL UNILATERAL LUMBAR ABLATION PROCEDURE, A BURN WAS NOTED AFTER REMOVING THE GROUNDING PAD. THE GROUNDING PAD HAD LEFT A CLEAR FILM ON THE PATIENT FOLLOWING REMOVAL AND A BLISTER WAS NOTED ON THE SKIN UNDER THE LEFT FLANK. THE PATIENTS SKIN WAS NOT PREPPED BEFORE PLACING THE GROUNDING PAD. THE PATIENT IS IN STABLE CONDITION.
Patient problems: Burn(s)
MAUDE report 8730589 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page