Atellica CH 930 Analyzer, Siemens Material Number 11067000… recallZ-0862-2020 · Siemens Healthcare Diagnostics, Inc. · initiated Oct 11, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
On the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and Chloride (Cl), the system will translate the decimal point separators to commas. This will result in invalid concentration values for the fluids, therefore the system will utilize default nominal values instead of lot specific IMT fluid concentration values for calculation of the calibrations and results for the Na, K, and Cl methods.
FDA root cause category: Under Investigation by firm
What the company is doing
The firm, Siemens Healthineers, sent an "Urgent Field Safety Notice ASW20-01.A.OUS" dated October, 2019 to its customers. The notice described the product, the problem and the actions to be taken. Actions to be Taken by the Customer 1. Scan the barcodes for IMT system fluids Std A and Std B for Na, K, and Cl at the system workstation using the manual barcode reader. Do not scan these barcodes at the module console. " When available, software v1.20.1 will be delivered as follows: " For systems running software v1.19.0 or higher, the software will be delivered electronically through the Siemens Smart Remote Service (SRS) and a yellow alert: "A new software update is available and is ready to install." will prompt the user to install the software. " For all other system configurations, you will be contacted by your local Siemens Customer Service representative to schedule the software installation. " Please review this letter with your Medical Director. " Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days via email to: [email protected]. " If you have received any complaints of illness or adverse events associated with the product listed in Table1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you any further questions, contact Quality Engineer, at 914-631-8000 or email: [email protected].
Recall details
- Recall number
- Z-0862-2020
- Status
- Terminated
- Initiated by the company
- Oct 11, 2019
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 1528
- Product code
- JGS
- FDA submissions
- K151767
- Distribution
- The products were distributed to the following foreign countries: Argentina, Belgium, Brazil, Denmark, Finland, France, Germany, Greece, Italy, Latvia, Netherlands, Poland, Portugal, Spain, Sweden, Switzerland, and Vietnam. No affected products distributed in the United States.
Product description
Atellica CH 930 Analyzer, Siemens Material Number 11067000, Software Versions V1.19.2 or 1.20.0 The Atellica CH 930 Analyzer is an automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens.
Code information (models, lots, serial numbers)
UDI: 00630414002163, Software Versions V1.19.2 or 1.20.0 when used with languages (countries) that use a comma delimiter instead of a decimal separator (i.e. 1,02 vs 1.02), AND when scanning barcodes at the module console.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls679 reports
Other devices under the same FDA submission
5 devices cleared under K151767 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Atellica CH Diluent 11099300Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall0 reports
- Atellica A-LYTE™ Integrated Multisensor Na+ / K+ / Cl- 11099315Siemens Healthcare Diagnostics Inc. · ELECTRODE, ION SPECIFIC, POTASSIUM25 reports
- Atellica Direct Load 11068008Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE3 reports
- Atellica CH Cleaner 11099303Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 report
- Atellica CH AlbP 11097530Siemens Healthcare Diagnostics Inc. · BROMCRESOL PURPLE DYE-BINDING, ALBUMIN1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page