STA UNICALIBRATOR (ref. 00675) recallZ-0737-2020 · Diagnostica Stago, Inc. · initiated Nov 18, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.

FDA root cause category: Process control

What the company is doing

On November 18, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail stating: FIELD SAFETY NOTICE Combined use of the following reagents STA UNICALIBRATOR (ref. 00675) and STA DEFICIENT VIII (ref. 00725) STA DEFICIENT IX (ref. 00724) Dear Customer, You are using the STA Unicalibrator (00675) and STA Deficient VIII (00725) and/or STA Deficient IX (00724) and according to our records, you have ordered and received in your laboratory one or several kits of the STA - Unicalibrator listed in Appendix 1. Following the identification of a defect, please see the information below about the reagents and lots listed in Appendix 1. Description: Following a customer complaint, Stago has investigated and confirmed a positive bias on some STA - Unicalibrator levels. These high levels affect tests using STA - Deficient IX and STA - Deficient VIII reagents. Stago has accordingly reduced the concerned Unicalibrator plasma levels (see new levels on the attached lot-specific barcode sheets). According to our internal testing results, the reduced Unicalibrator plasma levels result in a decrease of Factor VIII and IX results on normal and below-normal patient plasmas. This change has no impact on the classification of hemophiliac patients (type A or B) as severe, moderate or mild. The reassignment of the Unicalibrator plasma levels necessitated a change in the acceptance ranges on the corresponding Quality Control plasmas for Factor VIII and IX parameters with the STA - Deficient VIII and IX reagents. You will find new barcode sheets for each lot of the affected reagents in the attachments. Actions: Upon receipt of this letter, we kindly ask you to: - No longer use the values indicated for STA - Deficient VIII and IX reagents on the barcode sheets in your possession (for the reagents & lots listed in Appendix 1). - Use the new barcode sheets (attached): - For STA - Unicalibrator: Use the analyzer-spec

Recall details

Recall number
Z-0737-2020
Status
Terminated
Initiated by the company
Nov 18, 2019
Posted by the FDA
Dec 23, 2019
Recalling firm
Diagnostica Stago, Inc.
Quantity
5,498
Product code
GIZ
FDA submissions
K922021
Distribution
US: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK

Product description

STA UNICALIBRATOR (ref. 00675)

Code information (models, lots, serial numbers)

REF: 00675 Lot Numbers: 253180, 253964, 254721, 255108 UDIs: (01)13607450006756(11)171231(17)191231(10)253180(241)00675 (01)13607450006756(11)180630(17)200630(10)253964(241)00675 (01)13607450006756(11)181130(17)201130(10)254721(241)00675 (01)13607450006756(11)190228(17)210228(10)255108(241)00675