STA SYSTEM CONROL N + P (ref. 00678) recallZ-0738-2020 · Diagnostica Stago, Inc. · initiated Nov 18, 2019
Class II Terminated All recalls posted in December 2019
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.
FDA root cause category: Process control
What the company is doing
On November 18, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail stating: FIELD SAFETY NOTICE Combined use of the following reagents STA UNICALIBRATOR (ref. 00675) and STA DEFICIENT VIII (ref. 00725) STA DEFICIENT IX (ref. 00724) Dear Customer, You are using the STA Unicalibrator (00675) and STA Deficient VIII (00725) and/or STA Deficient IX (00724) and according to our records, you have ordered and received in your laboratory one or several kits of the STA - Unicalibrator listed in Appendix 1. Following the identification of a defect, please see the information below about the reagents and lots listed in Appendix 1. Description: Following a customer complaint, Stago has investigated and confirmed a positive bias on some STA - Unicalibrator levels. These high levels affect tests using STA - Deficient IX and STA - Deficient VIII reagents. Stago has accordingly reduced the concerned Unicalibrator plasma levels (see new levels on the attached lot-specific barcode sheets). According to our internal testing results, the reduced Unicalibrator plasma levels result in a decrease of Factor VIII and IX results on normal and below-normal patient plasmas. This change has no impact on the classification of hemophiliac patients (type A or B) as severe, moderate or mild. The reassignment of the Unicalibrator plasma levels necessitated a change in the acceptance ranges on the corresponding Quality Control plasmas for Factor VIII and IX parameters with the STA - Deficient VIII and IX reagents. You will find new barcode sheets for each lot of the affected reagents in the attachments. Actions: Upon receipt of this letter, we kindly ask you to: - No longer use the values indicated for STA - Deficient VIII and IX reagents on the barcode sheets in your possession (for the reagents & lots listed in Appendix 1). - Use the new barcode sheets (attached): - For STA - Unicalibrator: Use the analyzer-spec
Recall details
- Recall number
- Z-0738-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 18, 2019
- Posted by the FDA
- Dec 23, 2019
- Recalling firm
- Diagnostica Stago, Inc.
- Quantity
- 5,437
- Product code
- GGN
- FDA submissions
- K943518
- Distribution
- US: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK
Product description
STA SYSTEM CONROL N + P (ref. 00678)
Code information (models, lots, serial numbers)
REF: 00678 Lot Numbers: 253145, 253416, 253655, 254040, 254206, 254401, 254573, 254760 UDIs: (01)03607450006780(11)171130(17)191130(10)253145(241)00678 (01)03607450006780(11)180131(17)200131(10)253416(241)00678 (01)03607450006780(11)180331(17)200331(10)253655(241)00678 (01)03607450006780(11)180630(17)200630(10)254040(241)00678 (01)03607450006780(11)180731(17)200731(10)254206(241)00678 (01)03607450006780(11)180930(17)200930(10)254401(241)00678 (01)03607450006780(11)181031(17)201031(10)254573(241)00678 (01)03607450006780(11)181130(17)201130(10)254760(241)00678
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- STA® - System Control N + P 00678Diagnostica Stago · PLASMA, COAGULATION CONTROL1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 84402).
- Z-0737-2020 · TerminatedSTA UNICALIBRATOR (ref. 00675)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page