Skip to content

Affinity recallZ-0693-2020 · Medtronic Perfusion Systems · initiated Nov 14, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medtronic has identified an out of-specification condition exhibiting excess plastic (flash) in the arterial filter directly above the outlet port.

FDA root cause category: Process control

What the company is doing

Beginning 14-Nov-2019, Medtronic initiated an immediate Urgent Medical Device Recall via verbal communication to one (1) customer that was shipped the affected product according to Medtronic records. The customer was asked to identify and quarantine any unused affected product. The immediate communication was followed by a written urgent medical device recall notification beginning on 26-Nov-2019 to communicate the issue to the one (1) customer that received affected product. The consignee was asked to confirm receipt of urgent medical device recall notification via a Customer Confirmation Certificate. Medtronic is seeking the return of the affected product.

Recall details

Recall number
Z-0693-2020
Status
Terminated
Initiated by the company
Nov 14, 2019
Recalling firm
Medtronic Perfusion Systems
Quantity
24 units
Product code
DTM
FDA submissions
K952532
Distribution
CA

Product description

Medtronic Affinity, 38um Arterial Filter, Catalog number REF 351

Code information (models, lots, serial numbers)

UDI 20643169858835, Lot Numbers: 13316512

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page