AffinityRecalls & reported problems across all models
Medtronic, Inc. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 20
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 3 models of Affinity by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 20 adverse event reports in the FDA's MAUDE database, received between May 20, 2022 and Dec 29, 2025 — classified as 20 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (12 reports).
1 FDA recall record name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Affinity recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Nov 14, 2019Terminated
Medtronic has identified an out of-specification condition exhibiting excess plastic (flash) in the arterial filter directly above the outlet port.
Z-0693-2020 · Root cause: Process control · Recall details · FDA record ↗
Most reported Affinity problems
Device problems
- Fluid/Blood Leak12
- Crack3
- Material Deformation3
- Material Puncture/Hole3
- Air/Gas in Device1
- Fracture1
- Gas Output Problem1
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions19
- Hemorrhage/Blood Loss/Bleeding2
Affinity models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Affinity™ 61399401124Filter, blood, cardiopulmonary bypass, arterial line1 recall0 reports
- Affinity® 95214OXYGENATOR, CARDIOPULMONARY BYPASS19 reports
- Affinity® CB511OXYGENATOR, CARDIOPULMONARY BYPASS1 report
Recent Affinity adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionDec 29, 2025Fluid/Blood Leak; Gas Output Problem; Air/Gas in Device; Fluid/Blood Leak; Gas Output Problem; Air/Gas in Device
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF AN AFFINITY NT OXYGENATOR, IT WAS REPORTED THAT IT HAD BLOOD LEAKAGE AND PRODUCED FOAM, RESULTING IN DECREASED OXYGENATION EFFICIENCY. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 23904815 ↗MalfunctionMay 29, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING PRE-OPERATIVE PRIMING, THE INTERFACE OF THE AFFINITY NT OXYGENATOR WAS OBSERVED TO BE DAMAGED. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22103478 ↗MalfunctionMar 31, 2025Crack; Crack
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AFFINITY NT OXYGENATOR, IT WAS REPORTED THAT THE OXYGENATOR INTERFACE WAS DAMAGED. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21729188 ↗MalfunctionMar 7, 2025Material Deformation; Insufficient Device Problem Information; Material Deformation; Insufficient Device Problem Information
IT WAS REPORTED THAT PRIOR TO USE OF AN AFFINITY NT OXYGENATOR ONE INTERFACE WAS FLATTENED AND COULD NOT BE CONNECTED TO THE PIPELINE NORMALLY. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT AFTER OPENING THE PRODUCT PACKAGE THE DEVICE WAS OBSERVED TO BE DAMAGED. THE BOTTOM OF THE HEAT EXCHANGER INTERFACE WAS DAMAGED. THERE WAS NO LEAK OBSERVED ON THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21545462 ↗MalfunctionMar 5, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DURING USE OF AN AFFINITY NT OXYGENATOR THE DEVICE WAS LEAKING WHILE PERFORMING EXTRACORPOREAL CIRCULATION SUPPORT FOR THE PATIENT. THE DEVICE WAS REPLACED. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21523614 ↗MalfunctionJan 14, 2025Crack; Crack
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AFFINITY NT OXYGENATOR, IT WAS REPORTED THAT THE CONNECTION HAD BROKEN PIECES AT THE INTERFACE. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO THE PACKAGING (OUTER BOX, INNER PACKAGING OR STERILE BARRIER). NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER WERE DAMAGED. THERE WAS NO DAMAGE OBSERVED TO THE MAIN BODY OF THE OXYGENATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21151890 ↗MalfunctionNov 27, 2024Fracture; Material Deformation; Fracture; Material Deformation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THIS AFFINITY NT OXYGENATOR AND AFFINITY NT CARDIOTOMY VENOUS RESERVOIR, IT WAS REPORTED THAT LIQUID LEAKAGE OCCURRED DURING PRE-FILLING BEFORE SURGERY. IN THE SECOND EVENT, DEFORMATION OF THE CONSUMABLE INTERFACE WAS FOUND WHEN THE PACKAGE WAS OPENED BEFORE SURGERY. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20796512 ↗MalfunctionNov 27, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THIS AFFINITY NT OXYGENATOR AND AFFINITY NT CARDIOTOMY VENOUS RESERVOIR, IT WAS REPORTED THAT LIQUID LEAKAGE OCCURRED DURING PRE-FILLING BEFORE SURGERY. IN THE SECOND EVENT, DEFORMATION OF THE CONSUMABLE INTERFACE WAS FOUND WHEN THE PACKAGE WAS OPENED BEFORE SURGERY. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20796414 ↗MalfunctionMay 24, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF AN AFFINITY NT OXYGENATOR, THE CUSTOMER REPORTED A LEAK FROM THE TEMPERATURE PROBE. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19389718 ↗MalfunctionJan 30, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AFFINITY NT OXYGENATOR AND AFFINITY NT CARDIOTOMY VENOUS RESERVOIR, IT WAS REPORTED THAT THERE WAS LOSS. THE USE OF THE DEVICE WAS NOT SPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18604245 ↗MalfunctionJan 30, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AFFINITY NT OXYGENATOR AND AFFINITY NT CARDIOTOMY VENOUS RESERVOIR, IT WAS REPORTED THAT THERE WAS LOSS. THE USE OF THE DEVICE WAS NOT SPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18604232 ↗MalfunctionJan 29, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AFFINITY NT OXYGENATOR AND AFFINITY NT CARDIOTOMY VENOUS RESERVOIR, IT WAS REPORTED THAT THERE WAS A LOSS. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18594216 ↗MalfunctionOct 19, 2023Material Puncture/Hole; Material Puncture/Hole
MEDTRONIC RECEIVED INFORMATION THAT AT AN UNSPECIFIED TIME, THIS AFFINITY NT OXYGENATOR WAS CRACKED AND DAMAGED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WERE NO ADVERSE PATIENT EFFECTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17963443 ↗MalfunctionOct 4, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A AFFINITY NT OXYGENATOR AND AFFINITY NT CARDIOTOMY VENOUS RESERVOIR, IT WAS REPORTED THAT DURING AN EXTRACORPOREAL CIRCULATION PROCEDURE, THE CUSTOMER OBSERVED OOZING BLOOD AND THE DEVICE COULD NO LONGER BE USED. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE OXYGENATOR LEAKED BLOOD. THE CUSTOMER STATED THAT THE DEVICE ITSELF LEAKED. THE LOCATION OF THE LEAK WAS AT THE OXYGENATOR STORAGE TANK. THERE WAS NO BLOOD LOSS AS A RESULT OF THE LEAK. A TRANSFUSION WAS NOT REQUIRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO THE DEVICE, THE PACKAGE, OR THE CONTENTS WITHIN THE PACKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17867192 ↗MalfunctionSep 13, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A AFFINITY NT OXYGENATOR, IT WAS REPORTED THAT THE DEVICE WAS LEAKING BLOOD. THE D EVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO IMPACT TO THE PATIENT, THE SURGERY WENT WELL..
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17735301 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.