Dako Autostainer instruments: Autostainer Link 48, Model… recallZ-2074-2020 · Dako Denmark A/S · initiated Nov 5, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Dako Denmark A/S or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe tray, the updated Basic User Guide will reduce the potential risk of false negative results in the case of leakage of buffer and/or reagents that may impact staining.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The firm issued letters dated 10/30/2019 via email or standard mail, or phone call beginning 11/5/2019 notifying the customer of the issue, providing the risk to health, and actions that needed to be taken. On 09/03/2020, Agilent Dako sent Urgent Field Safety Notices to customers who were asked to take the following actions: 1) Carefully read and communicate to those within your organization of the updates stated in this notification letter and attached Autostainer Link Basic User Guide (rev. 06). Your local sales team will confirm with you that you have read and acknowledged this notification. 2) Your Agilent representative should be contacted immediately for assistance in the event of leakage. 3) Please discard previous versions of the Autostainer Link Basic User Guide that you are currently using.

Recall details

Recall number
Z-2074-2020
Status
Terminated
Initiated by the company
Nov 5, 2019
Posted by the FDA
May 21, 2020
Recalling firm
Dako Denmark A/S
Quantity
796 devices
Product code
KPA
Distribution
Distribution was nationwide. There was government/military distribution. Foreign distribution was made to Canada, Australia, Bangaladesh, Belarus, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Macedonia, Morocco, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Poland, Qatar, Romania, Russia, Saudi Arabia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Uruguay, and Vietnam.

Product description

Dako Autostainer instruments: Autostainer Link 48, Model AS480; Autostainer Plus Link Instrument, Model AS100; Autostainer, Model S3400; and Autostainer Plus, Model S3800.

Code information (models, lots, serial numbers)

All serial numbers of devices manufactured between 7/22/1997-6/28/2018 that do not have an installed syringe tray, UDI 05700572035497.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.