Medical device recalls: May 2020
The FDA posted 68 medical device recall records in May 2020, from 21 recall events by 19 companies. 7 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Siemens Medical Solutions USA, Inc · 2 events, 5 records
- Beckman Coulter Inc. · 2 events, 3 records
- Howmedica Osteonics Corp. · 1 event, 16 records
- Access Scientific LLC · 1 event, 12 records
- Osteomed, LLC · 1 event, 12 records
- Applied Medical Resources Corp · 1 event, 3 records, 3 Class I
- Philips North America, LLC · 1 event, 3 records
- Argon Medical Devices, Inc · 1 event, 2 records, 2 Class I
FDA root cause
- Packaging · 28
- Vendor change control · 12
- Process control · 6
- Nonconforming Material/Component · 4
- Process design · 3
- Device Design · 3
- Component change control · 3
- Package design/selection · 2
Class I recalls (7)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2039-2020 · Medtronic Vascular Galway DBA Medtronic Ireland · posted May 27, 2020 · Terminated
Increased risk of events associated with vascular complications (bleeding, hematoma, and dissection)
- Z-1995-2020 · Medtronic Navigation, Inc. · posted May 18, 2020 · Terminated
Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as…
- Z-1791-2020 · Argon Medical Devices, Inc · posted May 7, 2020 · Terminated
The stiffness of the soft end of the guidewire caused tissue perforation.
- Z-1792-2020 · Argon Medical Devices, Inc · posted May 7, 2020 · Terminated
The stiffness of the soft end of the guidewire caused tissue perforation.
- Z-1846-2020 · Applied Medical Resources Corp · posted May 7, 2020 · Terminated
The firm has received reports of tip separation during usage of the catheters, which may expose patients to unintended dislodgement of the tip from the body of the catheter.
- Z-1847-2020 · Applied Medical Resources Corp · posted May 7, 2020 · Terminated
The firm has received reports of tip separation during usage of the catheters, which may expose patients to unintended dislodgement of the tip from the body of the catheter.
- Z-1848-2020 · Applied Medical Resources Corp · posted May 7, 2020 · Terminated
The firm has received reports of tip separation during usage of the catheters, which may expose patients to unintended dislodgement of the tip from the body of the catheter.
Class II recalls (61)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2180-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2181-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2182-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2183-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2184-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2185-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2186-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2187-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2188-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2189-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2190-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2191-2020 · Osteomed, LLC · posted May 29, 2020 · Terminated
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
- Z-2136-2020 · Beckman Coulter Inc. · posted May 28, 2020 · Terminated
Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2…
- Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808.Class IIZ-2111-2020 · GE Healthcare, LLC · posted May 22, 2020 · Terminated
During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.
- Z-2069-2020 · Baylis Medical Company Inc. · posted May 21, 2020 · Terminated
There is a remote potential for the sterile packaging of lots of the transseptal guiding sheath to be compromised during shipping and handling. In the case that a packaging defect is not identified and the product is used in a patient, there may be a…
- NRG RF Transseptal Kit ↗Class IIZ-2070-2020 · Baylis Medical Company Inc. · posted May 21, 2020 · Terminated
There is a remote potential for the sterile packaging of the transseptal guiding sheath to be compromised during shipping and handling. In the case that a packaging defect is not identified and the product is used in a patient, there may be a remote risk…
- Z-2074-2020 · Dako Denmark A/S · posted May 21, 2020 · Terminated
The Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe tray, the updated Basic User Guide will reduce the potential risk of false…
- Z-2052-2020 · W L Gore & Associates, Inc. · posted May 20, 2020 · Terminated
The incorrect label was placed on both the primary and secondary package.
- Z-1972-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- BD PosiFlush SF Saline Flush Syringe 10 mL used in POWERWAND"XL Alternate Site Kit, 4Fr . REF/UDI… ↗Class IIZ-1973-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- Z-1974-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- Z-1975-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- Z-1976-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- Z-1977-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- Z-1978-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- Z-1979-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- Z-1980-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- BD PosiFlush SF Saline Flush Syringe 10 mL used in POWERWAND XL 3Fr 6cm Pediatric Kit. REF/UDI /… ↗Class IIZ-1981-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- Z-1982-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- Z-1983-2020 · Access Scientific LLC · posted May 14, 2020 · Terminated
Saline Flush Syringe compromised sterility due to holes in the packaging.
- Z-1971-2020 · Becton, Dickinson and Company, BD Biosciences · posted May 13, 2020 · Terminated
False positive reaction on the reagent.
- Z-2117-2020 · Abbott Gmbh & Co. KG · posted May 12, 2020 · Open, Classified
Abbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient…
- Z-1947-2020 · Sunrise Medical (US) LLC · posted May 11, 2020 · Terminated
Due to programming errors in the Controller, an increase in the set motor parameters can be made by the end user beyond tested safe limits.
- Z-1944-2020 · Beckman Coulter Inc. · posted May 9, 2020 · Terminated
A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke…
- Z-1945-2020 · Beckman Coulter Inc. · posted May 9, 2020 · Terminated
A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke…
- Z-1928-2020 · Edwards Lifesciences, LLC · posted May 8, 2020 · Terminated
Intermittent communication due to inconsistent crimping of connectors onto cables.
- Z-2084-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2085-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2086-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2087-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2088-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2089-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2090-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2091-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2092-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2093-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2094-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2095-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2096-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2097-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2098-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-2099-2020 · Howmedica Osteonics Corp. · posted May 7, 2020 · Terminated
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Z-1921-2020 · Philips North America, LLC · posted May 6, 2020 · Terminated
The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.
- Z-1922-2020 · Philips North America, LLC · posted May 6, 2020 · Terminated
The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.
- Z-1923-2020 · Philips North America, LLC · posted May 6, 2020 · Terminated
The bolt affixing the cooling unit to the rotor was not tightened to specification on the Philips Brilliance iCT systems.
- Z-1851-2020 · Nobel Biocare Usa Llc · posted May 4, 2020 · Terminated
The top label of article no. 37965, lot 12129104 (immediate packaging of endosseous dental implants ) may incorrectly indicate an implant length of 7mm, which is shorter than the actual length of the implant (10mm).
- Z-1957-2020 · Siemens Medical Solutions USA, Inc · posted May 4, 2020 · Terminated
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe…
- Z-1958-2020 · Siemens Medical Solutions USA, Inc · posted May 4, 2020 · Terminated
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe…
- Z-1959-2020 · Siemens Medical Solutions USA, Inc · posted May 4, 2020 · Terminated
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe…
- Z-2127-2020 · Siemens Medical Solutions USA, Inc · posted May 4, 2020 · Terminated
The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the…
- Z-2128-2020 · Siemens Medical Solutions USA, Inc · posted May 4, 2020 · Terminated
The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.