SnapLink recallZ-1257-2020 · Ormco/Sybronendo · initiated Jan 17, 2020

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ormco/Sybronendo or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The action is being taken due to an increase in complaints relating to the lower buccal tube slide doors not staying close. In the event of functional failure, broken/damage buccal tubes it may cause the tooth to rotate, possibly delay treatment and inconvenience the patient and clinician. If not noticed it may extend treatment by approximately 1 -3 months.

FDA root cause category: Device Design

What the company is doing

ISSUE: Ormco received an increase in complaints relating to lower Snaplink buccal tube slide doors not staying closed. We received 140 complaints from 2015 to Sep 2019, with no reported patient injuries. The incidence of this issue appears to be low. Approximately <0.1% of the Snaplink Buccal Tubes experience this issue due to an interaction between the Snaplink slide and the pin that enables the door to stay open or closed. RISK: In the event of functional failure, broken/damaged Snaplink may cause the tooth to rotate, possibly delay treatment and inconvenience the patient and clinician. If a patient or clinician does not notice this issue, it may extend treatment by approximately 1-3 months, depending on the patients visit schedule. If you experience this failure, please follow the recommendations below. RECOMMENDATION: Ormco is not requesting removal of Snaplink, from your facility. You can continue to use the product in your inventory. If you experience Snaplink buccal tube slide doors not staying closed, Ormco requests you to: (1) Call Ormco customer service at 1-800-854-1741 or email [email protected] to log any complaints related to this issue. (2) For patients who had Snaplink lowers placed, check the patient records and determine if you need an additional patient visit to check the integrity of the door to ensure it remains closed. (3) Instruct patients to check bracket after each brush to ensure bracket door remains closed. If door appears open, instruct patients to contact your office to schedule an appointment. (4) If the bracket door does not stay closed, use a double over tie, per typical orthodontic practices (Refer to pictures below) or remove the Snaplink and replace it with a new device. If the Snaplink is removed, please contact customer care (1-800-854-1741) for the product to be returned. (5) Complete the enclosed acknowledgement form and email the form to [email protected] within 10 business days. If

Recall details

Recall number
Z-1257-2020
Status
Terminated
Initiated by the company
Jan 17, 2020
Posted by the FDA
Feb 6, 2020
Recalling firm
Ormco/Sybronendo
Quantity
3,632,493 brackets
Product code
DZD
Distribution
US:AK,AL,AR,AZ,CA,CO,CN,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI,WV,WY OUS: Argentina, Australia, Barbados, Bahamas, Bolivia, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Hong Kong, Indonesia, Japan, Cambodia, Korea, Cayman Islands, Sri Lanka, Myanmar, Mexico, Malaysia, Netherlands, Nepal, Panama, Philippines, Paraguay, Russian Federation, Singapore, Thailand, Taiwan, Uruguay, Viet Nam, South Africa

Product description

SnapLink, Ref Part No. 438-2190, L6R SnapLink, Slot .022, TQ--28, Ang +2, Rot +2, Qty 10, Made in Mexico, CE 0086, Rx Only

Code information (models, lots, serial numbers)

All Lots UDI #s: 00889989002652 & 00889989002669

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 84538).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page