Medical device recalls: February 2020
The FDA posted 157 medical device recall records in February 2020, from 48 recall events by 36 companies. 40 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Theralase Inc. · 4 events, 4 records
- Implant Direct Sybron Manufacturing LLC · 3 events, 3 records
- Covidien Llc · 2 events, 29 records
- Deroyal Industries, Inc. Lafollette · 2 events, 11 records
- Teleflex Medical Inc · 2 events, 6 records, 4 Class I
- Ethicon, Inc. · 2 events, 4 records
- Abbott Point Of Care Inc. · 2 events, 3 records
- Siemens Healthcare Diagnostics, Inc. · 2 events, 3 records
FDA root cause
- Equipment maintenance · 28
- Packaging process control · 25
- Software design · 17
- Process change control · 17
- Process control · 12
- Device Design · 12
- Under Investigation by firm · 11
- Environmental control · 10
Class I recalls (40)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0942-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0943-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0944-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0945-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0946-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0947-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0948-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0949-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0950-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0951-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0952-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0953-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-0954-2020 · Philips Respironics, Inc. · posted Feb 15, 2020 · Terminated
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing…
- Z-1128-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1129-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1130-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1131-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1132-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1133-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1134-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1135-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1136-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1137-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1138-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1139-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1140-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- NC TRAVELER Coronary Dilatation Catheter, REF 1013157-12, Rx Only,NC Traveler RX 4.0 X 12MM ↗Class IZ-1141-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- NC TRAVELER Coronary Dilatation Catheter, REF 1013157-15, Rx Only,NC Traveler RX 4.0 X 15MM ↗Class IZ-1142-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-1143-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- NC TRAVELER Coronary Dilatation Catheter, REF 1013158-12, Rx Only,NC Traveler RX 4.5 X 12MM ↗Class IZ-1144-2020 · Abbott Vascular · posted Feb 14, 2020 · Terminated
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
- Z-0930-2020 · Tytek Medical Inc · posted Feb 12, 2020 · Terminated
A defect involving an occluded needle was discovered during a training exercise.
- Z-0955-2020 · Medtronic Inc. · posted Feb 7, 2020 · Open, Classified
There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump,…
- Z-0956-2020 · Medtronic Inc. · posted Feb 7, 2020 · Open, Classified
There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump,…
- Medtronic MiniMed 620G Insulin Infusion Pump, Ref #s/Model(s) - MMT-1510, MMT-1710, and MMT-1750 ↗Class IZ-0957-2020 · Medtronic Inc. · posted Feb 7, 2020 · Open, Classified
There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump,…
- Z-0958-2020 · Medtronic Inc. · posted Feb 7, 2020 · Open, Classified
There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump,…
- Z-1112-2020 · Teleflex Medical Inc · posted Feb 7, 2020 · Terminated
There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
- Z-1113-2020 · Teleflex Medical Inc · posted Feb 7, 2020 · Terminated
There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
- Z-1114-2020 · Teleflex Medical Inc · posted Feb 7, 2020 · Terminated
There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
- Z-1115-2020 · Teleflex Medical Inc · posted Feb 7, 2020 · Terminated
There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
- Z-0916-2020 · King Systems Corp. dba Ambu, Inc. · posted Feb 4, 2020 · Terminated
Exhibiting a reversed image. Although the image may appear normal, the users actions will be reversed on the display for left and right directions.
Class II recalls (111)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1477-2020 · Cordis Corporation · posted Feb 28, 2020 · Terminated
Cordis is recalling one (1) lot of POWERFLEX P3 PTA Dilatation Catheter due to the potential for body/shaft voids in the proximal seal area.
- Precision Xtra Blood Glucose & Ketone Monitoring System, UDI: 00093815715023, 00093815988144Class IIZ-1556-2020 · Abbott Diabetes Care, Inc. · posted Feb 28, 2020 · Terminated
Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate…
- Z-1557-2020 · Abbott Diabetes Care, Inc. · posted Feb 28, 2020 · Terminated
Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate…
- Z-1558-2020 · Abbott Diabetes Care, Inc. · posted Feb 28, 2020 · Terminated
Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate…
- Z-1599-2020 · Siemens Healthcare Diagnostics, Inc. · posted Feb 27, 2020 · Terminated
The assay may exhibit a positive bias in QC and patient results within the 10-day pack calibration interval.
- Z-1446-2020 · Covidien Llc · posted Feb 25, 2020 · Terminated
Sterile packaging may potentially be compromised, and use of products may result in increased risk for infection.
- Z-1447-2020 · Covidien Llc · posted Feb 25, 2020 · Terminated
Sterile packaging may potentially be compromised, and use of products may result in increased risk for infection.
- Z-1349-2020 · Randox Laboratories, Limited · posted Feb 24, 2020 · Terminated
Software version UI2550642107 for the RX Daytona + instrument released to correct the following issues: 1) The order of samples displayed is not sequential according to SID (sample identification number) when searching the run/results screen; 2) Incorrect…
- Z-1458-2020 · Nextremity Solutions · posted Feb 24, 2020 · Terminated
Proximal end of the driver in the affected product may be oversized in varying degrees, potentially making it difficult to connect with and remove from the proximal Nextra implant
- Z-1513-2020 · Siemens Healthcare Diagnostics, Inc. · posted Feb 24, 2020 · Terminated
Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as pre-menopausal when truly post- menopausal. This may lead to delayed…
- Z-1514-2020 · Siemens Healthcare Diagnostics, Inc. · posted Feb 24, 2020 · Terminated
Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as pre-menopausal when truly post- menopausal. This may lead to delayed…
- Z-1328-2020 · Philips Medical Systems Gmbh, DMC · posted Feb 20, 2020 · Terminated
When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of…
- Z-1329-2020 · Philips Medical Systems Gmbh, DMC · posted Feb 20, 2020 · Terminated
When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of…
- Z-1438-2020 · Polymer Technology Systems, Inc. · posted Feb 20, 2020 · Terminated
The firm is removing the product from the market after discussions with the FDA due to lack of 510(k) clearance.
- Z-1274-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1275-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1276-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1277-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1278-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1279-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1280-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1281-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1282-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1283-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1284-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1285-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1286-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1287-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1288-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1289-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1290-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1291-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1292-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1293-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1294-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1295-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1296-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1297-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1298-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1299-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1300-2020 · Covidien LLC · posted Feb 19, 2020 · Terminated
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
- Z-1302-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1303-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1304-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1305-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1306-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1307-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1308-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1309-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1310-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1311-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1312-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1313-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1314-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1315-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1316-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1317-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1318-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1319-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1320-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1321-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1322-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1323-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1324-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1325-2020 · Nobel Biocare Usa Llc · posted Feb 19, 2020 · Terminated
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
- Z-1410-2020 · Deroyal Industries, Inc. Lafollette · posted Feb 19, 2020 · Terminated
Cardinal Health gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1411-2020 · Deroyal Industries, Inc. Lafollette · posted Feb 19, 2020 · Terminated
Cardinal Health gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1412-2020 · Deroyal Industries, Inc. Lafollette · posted Feb 19, 2020 · Terminated
Cardinal Health gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1413-2020 · Deroyal Industries, Inc. Lafollette · posted Feb 19, 2020 · Terminated
Cardinal Health gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1414-2020 · Deroyal Industries, Inc. Lafollette · posted Feb 19, 2020 · Terminated
Cardinal Health gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1415-2020 · Deroyal Industries, Inc. Lafollette · posted Feb 19, 2020 · Terminated
Cardinal Health gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1416-2020 · Deroyal Industries, Inc. Lafollette · posted Feb 19, 2020 · Terminated
Cardinal Health gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1417-2020 · Deroyal Industries, Inc. Lafollette · posted Feb 19, 2020 · Terminated
Cardinal Health gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1418-2020 · Deroyal Industries, Inc. Lafollette · posted Feb 19, 2020 · Terminated
Cardinal Health gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1419-2020 · Deroyal Industries, Inc. Lafollette · posted Feb 19, 2020 · Terminated
Cardinal Health gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
- Z-1483-2020 · Deroyal Industries, Inc. Lafollette · posted Feb 19, 2020 · Terminated
The sterile packaging has the potential to be compromised.
- Z-1256-2020 · Edan Diagnostics · posted Feb 18, 2020 · Terminated
Firm became aware that it distributed 7 units in the U.S. that included a function that had not been cleared by U.S. FDA.
- Z-1262-2020 · Abbott Point Of Care Inc. · posted Feb 18, 2020 · Terminated
The I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and non-waived. Different CLIA standards apply to waived and non-waived tests as well as the…
- Z-1263-2020 · Abbott Point Of Care Inc. · posted Feb 18, 2020 · Terminated
The I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and non-waived. Different CLIA standards apply to waived and non-waived tests as well as the…
- Z-1265-2020 · Abbott Point Of Care Inc. · posted Feb 18, 2020 · Terminated
Marketed without a 510(k). Abbott Point of Care has determined that it does not have sufficient clinical evidence to fully characterize the performance of the blue i-STAT G3+ cartridges and therefore will no longer distribute the i-STAT G3+ (blue)…
- Z-1343-2020 · Hologic, Inc. · posted Feb 18, 2020 · Terminated
The recalling firm has received reports from the field of the C-arm on the Selenia Dimensions system seizing or dropping during field repairs and cleaning of the systems.
- Z-1475-2020 · Maquet Cardiovascular Us Sales, Llc · posted Feb 18, 2020 · Terminated
Potential leakage that may occur due to ruptured spot welds in the chamber, could generate a situation for the user such as risk for slipping or infection from contaminated water
- Z-1449-2020 · Medtronic CoreValve LLC · posted Feb 14, 2020 · Open, Classified
Demonstration Simulus Reusable Sizer Accessory Kits were being incorrectly sourced to customers. The demonstration product (product used within a wet lab) could have come in contact with non-human (porcine) tissue.
- MedtronicClass IIZ-1450-2020 · Medtronic CoreValve LLC · posted Feb 14, 2020 · Open, Classified
Demonstration Simulus Reusable Sizer Accessory Kits were being incorrectly sourced to customers. The demonstration product (product used within a wet lab) could have come in contact with non-human (porcine) tissue.
- Z-1169-2020 · DePuy Orthopaedics, Inc. · posted Feb 13, 2020 · Terminated
The device will not connect to the Specialist 2 Locating Outrigger which renders them unusable in surgery.
- MedlineClass IIZ-1357-2020 · Medline Industries Inc · posted Feb 11, 2020 · Terminated
The sterile pack inside the Complete Delivery System (CDS) did not go through a complete sterilization cycle, therefore the sterility of these components cannot be guaranteed.
- MedlineClass IIZ-1358-2020 · Medline Industries Inc · posted Feb 11, 2020 · Terminated
The sterile pack inside the Complete Delivery System (CDS) did not go through a complete sterilization cycle, therefore the sterility of these components cannot be guaranteed.
- Z-1365-2020 · BIOMERIEUX - Grenoble · posted Feb 11, 2020 · Terminated
Following a customer complaint about colored eluates, leading to invalid results bioMerieux has confirmed an issue with NUCLISENS Lysis Buffer, ref. 200292, lot 19040201. The issue is linked to drift of pH electrode dysfunction.
- Z-1420-2020 · Tosoh Bioscience Inc · posted Feb 11, 2020 · Terminated
A display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not…
- Z-1120-2020 · Flowonix Medical Inc · posted Feb 10, 2020 · Open, Classified
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve…
- Z-1121-2020 · Flowonix Medical Inc · posted Feb 10, 2020 · Open, Classified
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve…
- Z-1122-2020 · Flowonix Medical Inc · posted Feb 10, 2020 · Open, Classified
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve…
- Z-1123-2020 · Flowonix Medical Inc · posted Feb 10, 2020 · Open, Classified
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve…
- Z-1124-2020 · Flowonix Medical Inc · posted Feb 10, 2020 · Open, Classified
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve…
- Z-1125-2020 · Flowonix Medical Inc · posted Feb 10, 2020 · Open, Classified
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve…
- Z-1126-2020 · Implant Direct Sybron Manufacturing LLC · posted Feb 10, 2020 · Terminated
Due to a complaint received and internal investigation it was determined that implant vials were distributed with incorrect color coded vials and caps. The part/lot numbers were correct, the customer received the product they ordered, however the color on…
- Z-1145-2020 · Philips North America, LLC · posted Feb 10, 2020 · Terminated
Due to a software defect, when a digital subtraction angiography (DSA) is exported to the VesselNavigator application, the DSA is displayed without subtraction.
- Z-1408-2020 · Medtronic, PLC · posted Feb 10, 2020 · Open, Classified
Blades do not fit the McGRATH MAC video laryngoscope handle or were tight fitting on the handles which may result in a change in optical performance as displayed on the video screen
- Z-1116-2020 · Teleflex Medical Inc · posted Feb 7, 2020 · Terminated
The products are being recalled due to reports of breakage broke at the point of a welded joint either prior to use during setup or during use inside the patient's mouth.
- Z-1117-2020 · Teleflex Medical Inc · posted Feb 7, 2020 · Terminated
The products are being recalled due to reports of breakage broke at the point of a welded joint either prior to use during setup or during use inside the patient's mouth.
- Z-1257-2020 · Ormco/Sybronendo · posted Feb 6, 2020 · Terminated
The action is being taken due to an increase in complaints relating to the lower buccal tube slide doors not staying close. In the event of functional failure, broken/damage buccal tubes it may cause the tooth to rotate, possibly delay treatment and…
- Z-1258-2020 · Ormco/Sybronendo · posted Feb 6, 2020 · Terminated
The action is being taken due to an increase in complaints relating to the lower buccal tube slide doors not staying close. In the event of functional failure, broken/damage buccal tubes it may cause the tooth to rotate, possibly delay treatment and…
- Z-1334-2020 · Ethicon, Inc. · posted Feb 6, 2020 · Terminated
The product do not meet certain internal testing specifications.
- Z-1335-2020 · Ethicon, Inc. · posted Feb 6, 2020 · Terminated
The product do not meet certain internal testing specifications.
- Z-1336-2020 · Ethicon, Inc. · posted Feb 6, 2020 · Terminated
The product do not meet certain internal testing specifications.
- Z-1157-2020 · Implant Direct Sybron Manufacturing LLC · posted Feb 5, 2020 · Terminated
Through a customer complaint and and internal investigation it was determined that some implant packages contain the incorrect cover screw and extender. This could result in delay in surgery and/or a potential new surgical procedures to place the…
- Z-1330-2020 · Mindray DS USA, Inc. dba Mindray North America · posted Feb 5, 2020 · Terminated
The DP-30 displays an incorrect needle-guide bracket when used with the 65EC10EA model transducer.
- Z-0989-2020 · Ethicon, Inc. · posted Feb 4, 2020 · Terminated
Some sutures in the lot may exhibit a small section of braid fraying, which can lead to reduced tensile strength at that point.
- Z-1041-2020 · Implant Direct Sybron Manufacturing LLC · posted Feb 4, 2020 · Terminated
Some dental implant packages contain incorrect implants, which could lead to surgery rescheduling.
- Z-1301-2020 · Cardinal Health 200, LLC · posted Feb 4, 2020 · Terminated
The kits contained recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise…
- Z-1235-2020 · Cardinal Health 200, LLC · posted Feb 3, 2020 · Terminated
The kits contained recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could…
Class III recalls (6)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1259-2020 · Advanced Bionics, LLC · posted Feb 18, 2020 · Terminated
Three materials used in the makeup of the implant were mistakenly left off the list of bio-compatible materials in the firm's previously approved submission to the Pharmaceuticals and Medical Devices Agency (PMDA) of Japan.
- Z-1448-2020 · Biofire Defense · posted Feb 18, 2020 · Terminated
The firm has identified that an incorrect instruction for use (IFU) revision was included in the qualitative, multiplexed, nucleic acid-based in vitro diagnostic test kit.
- Z-1266-2020 · Theralase Inc. · posted Feb 13, 2020 · Terminated
Laser Probe (Model Number: TLC-2001) Electromagnetic Interference ( EMI ) in excess of IEC-60601-1-2 (4th edition) standard
- Z-1326-2020 · Theralase Inc. · posted Feb 13, 2020 · Terminated
Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 4th edition specification) shipped versus Power supply (Model Number: TLC-2003) MENB1100A2403F01 (24 VDC, 4.2 A,…
- Z-1348-2020 · Theralase Inc. · posted Feb 13, 2020 · Terminated
Power Pack (component) was shipped with an internal battery pack tested to UN-38.3, UL-2054 and IEC-62133 (2nd Edition) standards; however, not tested to IEC-60601-1-2 (3rd edition) standards
- Z-1354-2020 · Theralase Inc. · posted Feb 13, 2020 · Terminated
905 nm laser diodes may be below rated power output.