SIGNA Premier Nuclear Magnetic Resonance Imaging System -… recallZ-0874-2020 · GE Healthcare, LLC · initiated Dec 23, 2019
Class II Terminated All recalls posted in January 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
FDA root cause category: Software design
What the company is doing
This issue will be resolved by sending an Urgent Medical Device Correction letter to Medical Facility staff (Director of Clinical/Radiology, Risk Manager/Hospital Administrator, Director of Biomedical Engineering) with affected systems. This letter will inform the customer of the issue and instruct the user to ensure the displayed system date is correct. The long-term solution is to correct affected systems with updated software.
Recall details
- Recall number
- Z-0874-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 23, 2019
- Posted by the FDA
- Jan 28, 2020
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 62 (30 US, 32 OUS) total devices
- Product code
- LNH
- FDA submissions
- K171128
- Distribution
- Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Product description
SIGNA Premier Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Code information (models, lots, serial numbers)
Mfg. Lot or Serial# System ID 33-4-1031 415723SHMR19 To be provided RTD01204 To be provided DPN88672
Other devices under the same FDA submission
1 device cleared under K171128 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA SIGNA PremierGE Medical Systems, L.L.C. · Coil, magnetic resonance, specialty3 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 84655).
- Z-0875-2020 · TerminatedSIGNA Architect Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise rati…
- Z-0876-2020 · TerminatedSIGNA Pioneer Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio,…
- Z-0877-2020 · TerminatedSIGNA Vibrant Nuclear Magnetic Resonance Imaging System, 1.5T Signa HDx, 1.5T Signa HDxt, Signa HDi - Product Usage: is a whole body magnetic resonance scanner designed to support …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page