Skip to content

Dimension Enzymatic Creatinine (EZCR) reagent, SMN 10471520 recallZ-1232-2020 · Siemens Healthcare Diagnostics, Inc. · initiated Dec 18, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens Healthcare Diagnostics initiated a customer communication for this issue via letter beginning December 18, 2019.

Recall details

Recall number
Z-1232-2020
Status
Terminated
Initiated by the company
Dec 18, 2019
Posted by the FDA
Jan 24, 2020
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
2,198 units
Product code
JFY
FDA submissions
K073055
Distribution
US Nationwide

Product description

Dimension Enzymatic Creatinine (EZCR) reagent, SMN 10471520

Code information (models, lots, serial numbers)

UDI Number: 00842768030512 Lots: All lots

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 84612).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page