Polysorb Braided Absorbable Suture 2-0 VIOLET 36" GS-21… recallZ-1283-2020 · Covidien LLC · initiated Dec 18, 2019
Class II Terminated All recalls posted in February 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.
FDA root cause category: Equipment maintenance
What the company is doing
On December 18, 2109, Medtronic issued Urgent Medical Device Recall and Return Form notices to customers via courier service. Customers were advised to take the following required actions: 1. Please immediately quarantine and discontinue use of the affected item code and lot listed within the notice. 2. Please return affected product as indicated within the notice. All unused products from the affected item codes and lots must be returned. 3. If you have distributed any of the Polysorb braided or Biosyn monofilament absorbable sutures listed on the recall notice, please promptly forward the information from the letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory.
Recall details
- Recall number
- Z-1283-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 18, 2019
- Posted by the FDA
- Feb 19, 2020
- Recalling firm
- Covidien LLC
- Quantity
- 139,938 units total
- Product code
- GAM
- FDA submissions
- K000037
- Distribution
- US Nationwide distribution.
Product description
Polysorb Braided Absorbable Suture 2-0 VIOLET 36" GS-21, Item Code CL923 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in cardiovascular or neural tissue.
Code information (models, lots, serial numbers)
1) Lot - A9H1189Y Expiration Date - 2024-07-31; 2) Lot - A9J0345Y Expiration Date - 2024-08-31
Other devices under the same FDA submission
99 devices cleared under K000037 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Biosyn SM-5637Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID147 reports
- Biosyn GM-122Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID113 reports
- Biosyn SM-5638Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID113 reports
- Biosyn SM-5627Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID101 reports
- Biosyn SM-5679GCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID78 reports
- Biosyn SM-691Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID64 reports
- Biosyn SM-822Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID54 reports
- Biosyn GM-875Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID49 reports
- Biosyn SM-5687Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID48 reports
- Biosyn GM-324Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID46 reports
- Biosyn SM-693Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID43 reports
- Biosyn SM-5628Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID40 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 84769).
- Z-1274-2020 · TerminatedPolysorb Braided Absorbable Suture 3/0 VIOLET 30" CV-24 (75CM), Item Code CL461 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, …
- Z-1275-2020 · TerminatedPolysorb Braided Absorbable Suture 0 30" UNDYED HOS-11, Item Code CL537 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not …
- Z-1276-2020 · TerminatedPolysorb Braided Absorbable Suture 2/0 30 VIOLET GS-21, Item Code CL811 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not …
- Z-1277-2020 · TerminatedPolysorb Braided Absorbable Suture 3/0 30 VIOLET GS-11, Item Code CL815 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not …
- Z-1278-2020 · TerminatedPolysorb Braided Absorbable Suture 1 30 VIOLET GS-22, Item Code CL885 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in…
- Z-1279-2020 · TerminatedPolysorb Braided Absorbable Suture 13-0 VIOLET 30" GS-23, Item Code CL892 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but no…
- Z-1280-2020 · TerminatedPolysorb Braided Absorbable Suture 2 36" VIOLET GS-24, Item Code CL911 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not i…
- Z-1281-2020 · TerminatedPolysorb Braided Absorbable Suture 0 36 VIOLET GS-24, Item Code CL914 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in…
- Z-1282-2020 · TerminatedPolysorb Braided Absorbable Suture 1 36 VIOLET GS-24, Item Code CL915 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in…
- Z-1284-2020 · TerminatedPolysorb Braided Absorbable Suture 1 36 VIOLET GS-21, Item Code CL925 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in…
- Z-1285-2020 · TerminatedPolysorb Braided Absorbable Suture 2-0 UNDYED 36" GS-21, Item Code CL953 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not…
- Z-1286-2020 · TerminatedPolysorb Braided Absorbable Suture 2-0 VIOLET 30" V-20, Item Code GL123 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not …
- Z-1287-2020 · TerminatedPolysorb Braided Absorbable Suture 0 VIOLET 30" V-20, Item Code GL124 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in…
- Z-1288-2020 · TerminatedPolysorb Braided Absorbable Suture 2/0 UNDYED 30" V-30 (75CM), Item Code GL228 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, b…
- Z-1289-2020 · TerminatedPolysorb Braided Absorbable Suture 2-0 UNDYED 30" V-20, Item Code GL323 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not …
- Z-1290-2020 · TerminatedPolysorb Braided Absorbable Suture 3/0 30" UNDYED V-20, Item Code GLS322 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not…
- Z-1291-2020 · TerminatedPolysorb Braided Absorbable Suture 6/0 12"" VL SS-24 D/A, Item Code L1742K - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but n…
- Z-1292-2020 · TerminatedPolysorb Braided Absorbable Suture 6-0 18" UNDYED SS-24 D/A, Item Code L2752K - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, bu…
- Z-1293-2020 · TerminatedPolysorb Braided Absorbable Suture 6/0 18" UNDYED P-10, Item Code SL1625G - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but no…
- Z-1294-2020 · TerminatedPolysorb Braided Absorbable Suture UD 3/0 30 P-14, Item Code SL5679 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in c…
- Z-1295-2020 · TerminatedPolysorb Braided Absorbable Suture UD 4/0 18 P-13, Item Code SL5690 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in c…
- Z-1296-2020 · TerminatedPolysorb Braided Absorbable Suture 4/0 30 VIO SC-2 124K, Item Code SL654 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not…
- Z-1297-2020 · TerminatedPolysorb Braided Absorbable Suture 4/0 30 VIOLET CVF23, Item Code UL203 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not …
- Z-1298-2020 · TerminatedPolysorb Braided Absorbable Suture 0 30" VIOLET GU-46, Item Code UL877 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not i…
- Z-1299-2020 · TerminatedPolysorb Braided Absorbable Suture 2/0 30"VIOLET GU-46, Item Code UL878 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not …
- Z-1300-2020 · TerminatedBiosyn Monofilament Absorbable Suture 6/0 30" VIOLET CV-23, Item Code UM201
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page