Implant Direct, CE 0086, REF: 855710, Legacy 3 Impant… recallZ-1157-2020 · Implant Direct Sybron Manufacturing LLC · initiated Oct 11, 2019
Class II Terminated All recalls posted in February 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Implant Direct Sybron Manufacturing LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Through a customer complaint and and internal investigation it was determined that some implant packages contain the incorrect cover screw and extender. This could result in delay in surgery and/or a potential new surgical procedures to place the correct parts. The level of severity varies based on a one or two stage surgery.
FDA root cause category: Mixed-up of materials/components
What the company is doing
On October 11, 2019 the firm sent an "URGENT: MEDICAL DEVICE RECALL" letter via FedEx overnight priority mail. The letter advised the consignees of the recall. In addition, the letter states: We request that you take the following actions immediately:. 1. Please review your inventory for the affected product. If located, please quarantine the affected product and return the product to us according to the instructions on the attached form. 2. Please complete and return the Acknowledgement Form within fortyeight (48) hours of receipt of this notification, whether you have affected product or not. 3. If you are an authorized Implant Direct Sybron Manufacturing LLC distributor, we request that you identify those customers that may have been shipped the affected product lot and contact these customers to inform them of this issue within fortyeight (48) hours of receipt of this notification. 4. If you have any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425. Customer Care is available Monday through Friday, between the hours of 5am-5pm PST. If you or your patients have experienced any issues as a result of the affected products noted in this URGENT: MEDICAL DEVICE RECALL HHE 855710 100419 Rev 01 Status: RELEASED printed 10/10/2019 2:27:23 PM by Jose Trejo www.implantdirect.com | 888-649-6425 communication, you may voluntarily report the incident to the FDA through the MEDWATCH reporting system at the following: http://www.fda.gov/Safety/MedWatch/HowToReport/ucm053074.htm
Recall details
- Recall number
- Z-1157-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 11, 2019
- Posted by the FDA
- Feb 5, 2020
- Recalling firm
- Implant Direct Sybron Manufacturing LLC
- Quantity
- 78 implants
- Product code
- DZE
- FDA submissions
- K090234
- Distribution
- US: AR, AZ, CA, ID, IL, IN, KY, MD, ND, NH, NJ, NV, NY, Oh, OR, PA, TX, UT, WA, WI, WV, GA,MO
Product description
Implant Direct, CE 0086, REF: 855710, Legacy 3 Impant, Surface: SBM, Qty:1, Size:5.7mmD x 10mmL, 5.7mmD Platform, RxOnly, Sterile R,
Code information (models, lots, serial numbers)
Lot # 68209
Other devices under the same FDA submission
123 devices cleared under K090234 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Legacy™3 Implant 853710Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form1,106 reports
- Legacy™3 Implant 854710Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form815 reports
- Legacy™3 Implant 853711Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form758 reports
- Legacy™3 Implant 854210Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form715 reports
- Legacy™3 Implant 853210Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form505 reports
- Legacy™3 Implant 854211Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form475 reports
- Legacy™3 Implant 853713Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form461 reports
- Legacy™3 Implant 854708Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form428 reports
- Legacy™3 Implant 853211Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form407 reports
- Legacy™3 Implant 854711Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form403 reports
- Legacy™3 Implant 853708Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form378 reports
- Legacy™3 Implant 855210Implant Direct Sybron Manufacturing LLC · Implant, Endosseous, Root-Form340 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.