NC TREK recallZ-1129-2020 · Abbott Vascular · initiated Jan 29, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Vascular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Coronary Dilatation Catheters may exhibit difficulty or inability to deflate the balloon.
FDA root cause category: Process change control
What the company is doing
On 01/29/2020, the firm field personnel and Affiliates directly contacted the first customers beyond the firm's control via phone, email, or personal visit to facilitate rapid notification. The "URGENT MEDICAL DEVICE RECALL" Notification informed customers that specific lots of its Coronary Dilatation Catheters with diameters of 4.0 mm, 4.5 mm, and 5.0 mm may exhibit difficulty or inability to deflate the balloon due to weaker material proximal to the balloon bond resulting from excess heat exposure during manufacturing, and the potential risks with the use of the affected products include air embolism, thrombosis, myocardial infarction and additional intervention. Healthcare Professionals will be instructed to: " Reference the attached list of affected part numbers and lot numbers " Immediately stop using affected devices from these lots " Review your inventory, complete and return the attached Effectiveness Check Form " Return all unused affected product to Abbott " Share this notification with other relevant personnel in their organization The action the Recalling Firm is taking: * Stopped shipping affected lots * Will implement appropriate corrective actions to ensure product performance * Field Representatives can assist in identifying and returning affected devices * Will work with customers to replace returned units with similar devices, pending availability. For any questions, contact Customer Service Department at (800) 227-9902.
Recall details
- Recall number
- Z-1129-2020
- Status
- Terminated
- Initiated by the company
- Jan 29, 2020
- Posted by the FDA
- Feb 14, 2020
- Recalling firm
- Abbott Vascular
- Quantity
- Total 39687 NC TREK units (13891 US and 25796 OUS)
- Product code
- LOX
- FDA submissions
- K103153, K110134
- Distribution
- Worldwide Distribution: US (Nation-wide) and OUS (countries of): Algeria Jordan Argentina Korea Armenia Kuwait Aruba Lebanon Australia Malaysia Austria, Mauritius Bahrain Mexico Bangladesh Morocco Belgium Netherlands Brazil New Zealand Brunei Norway Bulgaria Oman Canada Pakistan Chile Panama China Peru Colombia Poland Costa Rica Portugal Croatia R¿union Cyprus Romania Czech Republic Russian Federation Denmark Saudi Arabia Dominican Republic Singapore Egypt Slovakia Estonia Slovenia Finland South Africa France Spain Germany Sri Lanka Greece State of Palestine Hong Kong Sweden Hungary Switzerland Iceland Taiwan, Province of China India Thailand Indonesia Tunisia Iran Turkey Iraq Ukraine Ireland United Arab Emirates Israel United Kingdom Italy Uruguay Japan Viet Nam
Product description
NC TREK Coronary Dilatation Catheter, REF 1012453-12 Rx Only, NC TREK RX 4.00 X 12MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion b) balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction c) balloon dilatation of a stent after implantation (balloon models 2.00 mm 5.00 mm only)
Code information (models, lots, serial numbers)
Device Identifier/GTIN: 08717648152061 Part Number: 1012453-12 Lot Numbers: 90730G1 90919G1 90731G1 90927G1 90818G1 91001G1 90822G1 91015G1 90905G1 91025G1 90918G1 91101G1
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 84826).
- Z-1128-2020 · TerminatedNC TREK Coronary Dilatation Catheter, REF 1012453-08 Rx Only, NC TREK RX 4.00 X 8MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the…
- Z-1130-2020 · TerminatedNC TREK Coronary Dilatation Catheter, REF 1012453-15 Rx Only, NC TREK RX 4.00 X 15MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of th…
- Z-1131-2020 · TerminatedNC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.00 X 20MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of th…
- Z-1132-2020 · TerminatedNC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the…
- Z-1133-2020 · TerminatedNC TREK Coronary Dilatation Catheter, REF 1012453-20 Rx Only, NC TREK RX 4.50 X 8MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the …
- Z-1134-2020 · TerminatedNC TREK Coronary Dilatation Catheter, REF 1012454-15 Rx Only, NC TREK RX 4.50 X 15MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of th…
- Z-1135-2020 · TerminatedNC TREK Coronary Dilatation Catheter, REF 1012454-20 Rx Only, NC TREK RX 4.50 X 20MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of th…
- Z-1136-2020 · TerminatedNC TREK Coronary Dilatation Catheter, REF 1012455-08 Rx Only, NC TREK RX 5.00 X 8MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the…
- Z-1137-2020 · TerminatedNC TREK Coronary Dilatation Catheter, REF 1012455-12 Rx Only, NC TREK RX 5.00 X 12MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of th…
- Z-1138-2020 · TerminatedNC TREK Coronary Dilatation Catheter, REF 1012455-15 Rx Only,NC TREK RX 5.00 X 15MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of the…
- Z-1139-2020 · TerminatedNC TREK Coronary Dilatation Catheter, REF 1012455-120 Rx Only,NC TREK RX 5.00 X 20MM BDC The NC TREK RX Coronary Dilatation Catheter is indicated for: a) balloon dilatation of th…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page