Precision Xtra Blood Glucose & Ketone Monitoring System… recallZ-1556-2020 · Abbott Diabetes Care, Inc. · initiated Nov 14, 2019
Class II Terminated All recalls posted in February 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Diabetes Care, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate meter disinfection or improper handling of contaminated meters, which may cause infection or exacerbation of an existing infection.
FDA root cause category: Labeling False and Misleading
What the company is doing
On 11/14/19, Urgent Medical Device Correction notices were mailed to customers via overnight delivery. Customers were told to continue to use instructions for use regarding the cleaning of meters for use on single patients only. Customers were further asked to complete and return customer response forms. Customers with additional questions were encouraged to call customer service in 1-888-522-5226.
Recall details
- Recall number
- Z-1556-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 14, 2019
- Posted by the FDA
- Feb 28, 2020
- Recalling firm
- Abbott Diabetes Care, Inc.
- Product code
- NBW
- FDA submissions
- K040814
- Distribution
- U.S.: KY, NY, TN, MA, IL, SC, VT, FL, CA, NC, MS, NJ, TX, AL, GA, OK, CT, PA, NE, ND, MT, MO, DC, ME, LA, KS, ID, OH, MN, NM, VA, IN
Product description
Precision Xtra Blood Glucose & Ketone Monitoring System, UDI: 00093815715023, 00093815988144
Code information (models, lots, serial numbers)
Not associated with specific model/lot numbers.
Other devices under the same FDA submission
1 device cleared under K040814 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Precision Xtra 71502Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER12 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 84866).
- Z-1557-2020 · TerminatedOptium EZ Blood Glucose Monitoring System, UDI: 00093815710653
- Z-1558-2020 · TerminatedFreeStyle Precision H Blood Glucose and Ketone Monitoring System, UDI: 00093815711810
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page