Elecsys CA 19-9, Cat. No. 07027028190 - Product Usage: the… recallZ-1444-2020 · Roche Diagnostics Operations, Inc. · initiated Feb 7, 2020

Class II Terminated All recalls posted in March 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm has received an increased number of complaints concerning non-reproducible elevated results ("high flyers") for one lot of Elecsys CA 19-9 reagent, which may affect clinical interpretation.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On February 7, 2020, the firm issued an Urgent Medical Device Correction letter to affected customers. Customers were informed of the product issue and given two options for CA 19-9 testing until another reagent lot is available for the cobas e 801 module: 1) Use a different Roche platform for testing, or 2) Implement a workaround, described in detail in the customer letter. The letter also provided instructions for receiving credit depending on which option was elected. Roche will notify customers when a new lot of CA 19-9 reagent for the cobas e 801 is available for ordering. ***Updated 10/5/20*** On 10/2/20, the firm issued an updated letter to customers to communicate an expansion to the recall. An additional lot was added to the recall scope.

Recall details

Recall number
Z-1444-2020
Recall class
Class II
Status
Terminated
Initiated by the company
Feb 7, 2020
Posted by the FDA
Mar 5, 2020
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
684
Product code
NIG
FDA submissions
K050231
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, FL, HI, IL, IN, KS, KY, MA, ME, MI, NC, NH, NJ, NY, TN, TX, WA and WI.

Product description

Elecsys CA 19-9, Cat. No. 07027028190 - Product Usage: the product is an immunoassay for monitoring and management of pancreatic cancer

Code information (models, lots, serial numbers)

Lot 416245 ***Updated 10/5/20*** 483123 ***Updated 8/5/22*** Additional released lots are subject to the same workaround: 46449, 50474,525510, and 568976

Other devices under the same FDA submission

2 devices cleared under K050231 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.