Medical device recalls: March 2020
The FDA posted 50 medical device recall records in March 2020, from 21 recall events by 17 companies. 11 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Siemens Medical Solutions USA, Inc · 2 events, 4 records
- Boston Scientific Corporation · 2 events, 3 records, 1 Class I
- Butterfly Network, Inc. · 2 events, 2 records
- CME America, LLC · 2 events, 2 records, 2 Class I
- Medtronic MiniMed · 1 event, 16 records
- CareFusion 303, Inc. · 1 event, 4 records, 4 Class I
- Gentell, Inc · 1 event, 4 records
- Northgate Technologies, Inc. · 1 event, 3 records
FDA root cause
- Device Design · 24
- Nonconforming Material/Component · 8
- No Marketing Application · 6
- Under Investigation by firm · 4
- Software design · 4
- Process change control · 2
- Environmental control · 1
- Component design/selection · 1
Class I recalls (11)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1555-2020 · Boston Scientific Corporation · posted Mar 25, 2020 · Open, Classified
Potential for tip detachment of Imager II 5F Angiographic Catheters
- Z-1474-2020 · CME America, LLC · posted Mar 18, 2020 · Terminated
Infusion Pump Systems may have a delivery inaccuracy of up to 13%, which may result in 1) faster than expected drug delivery when infusing at a very low rate (0.1 mL/h), or 2) slower than expected drug delivery when infusing at high flow rates…
- BodyGuard Microset REF A120-003XYVA, Sterile EO, Tubing: 150 in. (380 cm), Prime Vol: approx. 8mlClass IZ-1442-2020 · CME America, LLC · posted Mar 11, 2020 · Terminated
When using a specific tubing set with infusion pump system, may result in under deliver of fluids.
- Z-1427-2020 · Heartware, Inc. · posted Mar 9, 2020 · Open, Classified
Potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.
- Z-1428-2020 · Heartware, Inc. · posted Mar 9, 2020 · Open, Classified
Potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.
- Z-1527-2020 · Micro Therapeutics Inc, Dba Ev3 Neurovascular · posted Mar 9, 2020 · Terminated
Embolization device delivery system may fracture at the distal section during device implantation. A weakened bond may result in unintended separation, where the distal portion of the device delivery system remains in the patient. If a Pipeline Flex…
- Z-1528-2020 · Micro Therapeutics Inc, Dba Ev3 Neurovascular · posted Mar 9, 2020 · Terminated
Embolization device delivery system may fracture at the distal section during device implantation. A weakened bond may result in unintended separation, where the distal portion of the device delivery system remains in the patient. If a Pipeline Flex…
- Z-1359-2020 · CareFusion 303, Inc. · posted Mar 5, 2020 · Open, Classified
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result…
- Z-1360-2020 · CareFusion 303, Inc. · posted Mar 5, 2020 · Open, Classified
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result…
- Z-1361-2020 · CareFusion 303, Inc. · posted Mar 5, 2020 · Open, Classified
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result…
- Z-1362-2020 · CareFusion 303, Inc. · posted Mar 5, 2020 · Open, Classified
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result…
Class II recalls (39)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- C-Reactive Protein Kit for use on the SPAplus, REF NK044.S.A, Rx only, UDI (01)05051700018456, ↗Class IIZ-1619-2020 · The Binding Site Group, Ltd. · posted Mar 31, 2020 · Terminated
A reagent in C reactive protein kits is not performing to specification and samples with known concentrations less-than-or-equal to 10 mg/L may be affected
- Z-1581-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1582-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1583-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1584-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1585-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1586-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1587-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1588-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1589-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1590-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1591-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1592-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1593-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1594-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1595-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1596-2020 · Medtronic MiniMed · posted Mar 26, 2020 · Open, Classified
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a…
- Z-1597-2020 · Elekta Impac Software · posted Mar 26, 2020 · Open, Classified
The user may inadvertently enter Metric values into Height and Weight fields labeled with US Standard Units.
- Z-1564-2020 · Capintec Inc · posted Mar 24, 2020 · Terminated
It is possible that the collimators have screw hole locations that were insufficiently tapped. There is a potential for the collimator assembly to fall and come into contact with a patient or operator.
- Probe, 9 FR AUTOLITHClass IIZ-1883-2020 · Northgate Technologies, Inc. · posted Mar 24, 2020 · Terminated
As part of an investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing.
- AUTOLITH URO-TCH 9FR PROBE 54CMClass IIZ-1884-2020 · Northgate Technologies, Inc. · posted Mar 24, 2020 · Terminated
Investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing.
- Z-1885-2020 · Northgate Technologies, Inc. · posted Mar 24, 2020 · Terminated
Investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing.
- Z-1600-2020 · Butterfly Network, Inc. · posted Mar 17, 2020 · Terminated
Distributed without an approved 510(k)
- Z-1601-2020 · Butterfly Network, Inc. · posted Mar 17, 2020 · Terminated
Distributed without an approved 510(k)
- Z-1495-2020 · Siemens Medical Solutions USA, Inc · posted Mar 13, 2020 · Terminated
A software issue could potentially cause the stand and table movements to be blocked.
- Z-1496-2020 · Siemens Medical Solutions USA, Inc · posted Mar 13, 2020 · Terminated
A software issue could potentially cause the stand and table movements to be blocked.
- Z-1497-2020 · Siemens Medical Solutions USA, Inc · posted Mar 13, 2020 · Terminated
A software issue could potentially cause the stand and table movements to be blocked.
- Z-1444-2020 · Roche Diagnostics Operations, Inc. · posted Mar 5, 2020 · Terminated
The firm has received an increased number of complaints concerning non-reproducible elevated results ("high flyers") for one lot of Elecsys CA 19-9 reagent, which may affect clinical interpretation.
- Z-1623-2020 · Gentell, Inc · posted Mar 5, 2020 · Open, Classified
During an FDA audit, it was discovered that the product was not properly registered with FDA.
- Z-1624-2020 · Gentell, Inc · posted Mar 5, 2020 · Open, Classified
During an FDA audit, it was discovered that the product was not properly registered with FDA.
- Z-1625-2020 · Gentell, Inc · posted Mar 5, 2020 · Open, Classified
During an FDA audit, it was discovered that the product was not properly registered with FDA.
- Z-1626-2020 · Gentell, Inc · posted Mar 5, 2020 · Open, Classified
During an FDA audit, it was discovered that the product was not properly registered with FDA.
- Ysio, model no. 10281013 ↗Class IIZ-1501-2020 · Siemens Medical Solutions USA, Inc · posted Mar 3, 2020 · Terminated
The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the systems…
- Z-1511-2020 · Boston Scientific Corporation · posted Mar 3, 2020 · Terminated
Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software applications that have not been approved…
- Z-1512-2020 · Boston Scientific Corporation · posted Mar 3, 2020 · Terminated
Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software applications that have not been approved…
- Z-1543-2020 · Biomet, Inc. · posted Mar 3, 2020 · Terminated
A 15mm Shoulder Mini Humeral Stem was misidentified as a 9mm Shoulder Mini Humeral Stem.
- Z-1147-2020 · Stryker Corporation · posted Mar 2, 2020 · Terminated
Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone…
- Z-1148-2020 · Stryker Corporation · posted Mar 2, 2020 · Terminated
Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone…
- Z-1421-2020 · Siemens Healthcare Diagnostics, Inc. · posted Mar 2, 2020 · Open, Classified
Some lots ending in "17", "18", "19", or higher may have cuvette defects allowing water from the water bath to contaminate the interior of the cuvette, which may falsely elevate or depress sample results.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.