Skip to content

The FastPack System Comlete Immunoassay Kit recallZ-1672-2020 · Qualigen Inc · initiated Nov 6, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Qualigen Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The chemiluminescent immunoassay has been noted to have higher than expected results which may result in inaccurate quantitative determinations of Sex Hormone Binding Globulin in human serum and plasma.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On October 31, 2019, a "Urgent: Product Recall" letter was sent to customer via E-Mail. In addition to informing consignees about the recall the customer notification asked consignees to take the following actions: 1. The firm will provide replacement kits at no cost as soon as they become available. 2. For any kits that have been purchased a refund can be requested in place of replacement kits. 3. As part of the product replacement process, please confirm that any remaining SHBG FastPacks/Kits of Lot Number 1908038-1P have been discarded on the form below. 4. The firm will call to confirm that you have received the letter and provide any additional details you may require. For questions, please contact System Support at [email protected].

Recall details

Recall number
Z-1672-2020
Status
Terminated
Initiated by the company
Nov 6, 2019
Recalling firm
Qualigen Inc
Quantity
24 Kits (30 test packs each) 720 tests
Product code
CDZ
FDA submissions
K182521
Distribution
US Nationwide distribution in the states of AL, AK, CA, HI, and TX.

Product description

The FastPack System Comlete Immunoassay Kit Immunoassay for the Quantitative Measurement of SHBG is a chemiluminescent immunoassay intended for the quantitative determination of Sex Hormone Binding Globulin in human serum and plasma on the FastPack System. The FastPack IP SHBG assay is intended for use as an aid in the diagnosis of androgen disorders. Each Kit Contains 30 FastPack IP SHBG Immunoassays 1 Calibrator 1 x 3 mL Vial 1 Control 1 1 x 3 mL Vial 1 Control 2 1 x 3 mL Vials 32 Sample Diluent 0.9 mL Vials REF 25000081 IVD CE 1908038-1P 2020-02-22 (01) 20816467020345 (17) 200222 (10) 1908038-1P

Code information (models, lots, serial numbers)

1908038-1P 2020-02-22 (01) 20816467020345 (17) 200222 (10) 1908038-1P

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page