Skip to content

FastPack® IP SHBG FastPack® IP SHBG 30 ImmunoassayRecalls & reported problems

Qualigen Inc. · 30 Test SHBG Immunoassay Complete Kit w/Calibrator and Controls · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 3, 2024Class IIIOpen, Classified

    Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.

    Z-0805-2025 · Root cause: Labeling Change Control · Recall details · FDA record ↗

  2. Nov 6, 2019Class IITerminated

    The chemiluminescent immunoassay has been noted to have higher than expected results which may result in inaccurate quantitative determinations of Sex Hormone Binding Globulin in human serum and plasma.

    Z-1672-2020 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Other Single (Specified) Analyte Controls (Assayed And Unassayed) devices

All Single (Specified) Analyte Controls (Assayed And Unassayed) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page