Pharos Excimer Laser, Model: EX-308 - Product Usage: The… recallZ-1631-2020 · Ra Medical Systems, Inc. · initiated Aug 23, 2019

Class II Terminated All recalls posted in April 2020

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ra Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device to fail to cease firing as long as the foot pedal or handpiece switch is depressed. This malfunction can cause overtreatment, which could result in strong erythema, and possibly cause blisters.

FDA root cause category: Software change control

What the company is doing

On 08/23/2019, Medical Device Correction was mailed to customers. Customers were informed that the recalling firm's service department would contact them to replace the software. Unitl the software is replaced, do not use doses less than 100 mJ. During service visits customers were informed that software was updated to 1.06 ZRPL

Recall details

Recall number
Z-1631-2020
Recall class
Class II
Status
Terminated
Initiated by the company
Aug 23, 2019
Posted by the FDA
Apr 1, 2020
Recalling firm
Ra Medical Systems, Inc.
Quantity
10
Product code
GEX
FDA submissions
K062963
Distribution
U.S. Nationwide distribution in the states of CA, FL, OR, MS, NY.

Product description

Pharos Excimer Laser, Model: EX-308 - Product Usage: The Pharos EX-308 Excimer Laser System is designed to treat the skin disorders psoriasis, vitiligo, atopic dermatitis, and leukoderma using phototherapy. The excimer laser system produces collimated 308 nm radiation, within the invisible region of the electromagnetic spectrum. The UV radiation falls within the range of ultraviolet B (UVB), which is used to treat photo-responsive skin disorders.

Code information (models, lots, serial numbers)

All serial numbers.

Other devices under the same FDA submission

1 device cleared under K062963 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.