Medical device recalls: April 2020
The FDA posted 118 medical device recall records in April 2020, from 22 recall events by 20 companies. 2 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Beckman Coulter Inc. · 2 events, 3 records
- Philips North America, LLC · 2 events, 3 records
- Biomet, Inc. · 1 event, 33 records
- K2M, Inc · 1 event, 24 records
- Boston Scientific · 1 event, 12 records
- Welch Allyn, Inc/Mortara · 1 event, 10 records
- Canon Medical System, USA, INC. · 1 event, 8 records
- Boston Scientific Neuromodulation Corporation · 1 event, 5 records
FDA root cause
- Process control · 37
- Device Design · 35
- Component change control · 12
- Software design · 11
- Process design · 5
- Error in labeling · 5
- Radiation Control for Health and Safety Act · 4
- Nonconforming Material/Component · 2
Class I recalls (2)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Langston Dual Lumen Catheter 6F Model 5540. For delivery of contrast medium in angiographic studiesClass IZ-1747-2020 · Vascular Solutions, Inc. · posted Apr 27, 2020 · Terminated
The inner lumen of the Langston catheter may separate from the hub during pressure injection.
- Z-1567-2020 · LeMaitre Vascular, Inc. · posted Apr 1, 2020 · Open, Classified
There has been an increased trend in reports of catheters failing to deflate during use, which may ultimately lead to tip separation.
Class II recalls (116)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1809-2020 · Philips North America, LLC · posted Apr 29, 2020 · Terminated
When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the…
- Z-1810-2020 · Welch Allyn, Inc/Mortara · posted Apr 29, 2020 · Terminated
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Z-1811-2020 · Welch Allyn, Inc/Mortara · posted Apr 29, 2020 · Terminated
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Z-1812-2020 · Welch Allyn, Inc/Mortara · posted Apr 29, 2020 · Terminated
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Z-1813-2020 · Welch Allyn, Inc/Mortara · posted Apr 29, 2020 · Terminated
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Z-1814-2020 · Welch Allyn, Inc/Mortara · posted Apr 29, 2020 · Terminated
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Z-1815-2020 · Welch Allyn, Inc/Mortara · posted Apr 29, 2020 · Terminated
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Z-1816-2020 · Welch Allyn, Inc/Mortara · posted Apr 29, 2020 · Terminated
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Z-1817-2020 · Welch Allyn, Inc/Mortara · posted Apr 29, 2020 · Terminated
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Z-1818-2020 · Welch Allyn, Inc/Mortara · posted Apr 29, 2020 · Terminated
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Z-1819-2020 · Welch Allyn, Inc/Mortara · posted Apr 29, 2020 · Terminated
Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Z-1826-2020 · Elekta Inc · posted Apr 29, 2020 · Terminated
A faulty locking piece of the Instrument Carrier that does not fulfil the requirement of locking force to the Arc has been found during an internal check of a Leksell Vantage Stereotactic Arc System.
- Lunar iDXA bone densitometer ↗Class IIZ-1709-2020 · GE Healthcare, LLC · posted Apr 24, 2020 · Open, Classified
An issue with the Lunar iDXA control panel may cause the start button to not release and potentially contribute to an accidental radiation occurrence.
- Z-1710-2020 · GE Healthcare, LLC · posted Apr 24, 2020 · Open, Classified
An issue with the Lunar iDXA control panel may cause the start button to not release and potentially contribute to an accidental radiation occurrence.
- Z-1711-2020 · GE Healthcare, LLC · posted Apr 24, 2020 · Open, Classified
An issue with the Lunar iDXA control panel may cause the start button to not release and potentially contribute to an accidental radiation occurrence.
- Z-1712-2020 · GE Healthcare, LLC · posted Apr 24, 2020 · Open, Classified
An issue with the Lunar iDXA control panel may cause the start button to not release and potentially contribute to an accidental radiation occurrence.
- Z-2008-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2009-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2010-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2011-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2012-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2013-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2014-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2015-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2016-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2017-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2018-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2019-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2020-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2021-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2022-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2023-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2024-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2025-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2026-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2027-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2028-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2029-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2030-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-2031-2020 · K2M, Inc · posted Apr 24, 2020 · Terminated
The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.
- Z-1766-2020 · Philips North America, LLC · posted Apr 23, 2020 · Terminated
The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information…
- Z-1767-2020 · Philips North America, LLC · posted Apr 23, 2020 · Terminated
The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information…
- Z-1768-2020 · Acutus Medical Inc · posted Apr 23, 2020 · Terminated
It has been determined that the manufacturing process may have left foreign material particulates on the finished device. This could result in a range of event from procedure delay to pulmonary embolism.
- Z-1769-2020 · Acutus Medical Inc · posted Apr 23, 2020 · Terminated
It has been determined that the manufacturing process may have left foreign material particulates on the finished device. This could result in a range of event from procedure delay to pulmonary embolism.
- Z-1770-2020 · Acutus Medical Inc · posted Apr 23, 2020 · Terminated
It has been determined that the manufacturing process may have left foreign material particulates on the finished device. This could result in a range of event from procedure delay to pulmonary embolism.
- Z-1771-2020 · Acutus Medical Inc · posted Apr 23, 2020 · Terminated
It has been determined that the manufacturing process may have left foreign material particulates on the finished device. This could result in a range of event from procedure delay to pulmonary embolism.
- Z-1751-2020 · Beckman Coulter Inc. · posted Apr 20, 2020 · Terminated
Cross-reactivity testing may not meet the claims indicated in the Opiate 2000 ng (OP2) Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.
- Z-1752-2020 · Beckman Coulter Inc. · posted Apr 20, 2020 · Terminated
Cross-reactivity testing may not meet the claims indicated in the Opiate 300 ng (OP)Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.
- Z-1849-2020 · Epimed International · posted Apr 16, 2020 · Terminated
Incorrect expiration date on introduction cannula packaging.
- Z-1850-2020 · Epimed International · posted Apr 16, 2020 · Terminated
Incorrect expiration date on introduction cannula packaging.
- Portex LineDraw Arterial Blood Sampling Syringe with Dry Lithium Heparin for Gases and ElectrolytesClass IIZ-1715-2020 · Smiths Medical ASD, Inc. · posted Apr 15, 2020 · Terminated
Portex Pro-Vent and LineDraw Arterial Blood Sampling syringes may have broken luers.
- Portex ProVent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and ElectrolytesClass IIZ-1716-2020 · Smiths Medical ASD, Inc. · posted Apr 15, 2020 · Terminated
Lots of Smiths Medical Portex Pro-Vent and LineDraw Arterial Blood Sampling syringes may have broken luers.
- NavLock Tracker *** Rx OnlyClass IIZ-1693-2020 · Medtronic Navigation, Inc. · posted Apr 10, 2020 · Terminated
Product was incorrectly assembled which could affect navigation accuracy of the device.
- Z-1694-2020 · Boston Scientific Neuromodulation Corporation · posted Apr 10, 2020 · Terminated
The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.
- Z-1695-2020 · Boston Scientific Neuromodulation Corporation · posted Apr 10, 2020 · Terminated
The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.
- Z-1696-2020 · Boston Scientific Neuromodulation Corporation · posted Apr 10, 2020 · Terminated
The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.
- Z-1697-2020 · Boston Scientific Neuromodulation Corporation · posted Apr 10, 2020 · Terminated
The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.
- Z-1698-2020 · Boston Scientific Neuromodulation Corporation · posted Apr 10, 2020 · Terminated
The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.
- Z-1719-2020 · Canon Medical System, USA, INC. · posted Apr 10, 2020 · Open, Classified
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
- Z-1720-2020 · Canon Medical System, USA, INC. · posted Apr 10, 2020 · Open, Classified
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
- Z-1721-2020 · Canon Medical System, USA, INC. · posted Apr 10, 2020 · Open, Classified
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
- Z-1722-2020 · Canon Medical System, USA, INC. · posted Apr 10, 2020 · Open, Classified
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
- Z-1723-2020 · Canon Medical System, USA, INC. · posted Apr 10, 2020 · Open, Classified
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
- Z-1724-2020 · Canon Medical System, USA, INC. · posted Apr 10, 2020 · Open, Classified
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
- Z-1725-2020 · Canon Medical System, USA, INC. · posted Apr 10, 2020 · Open, Classified
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
- Z-1726-2020 · Canon Medical System, USA, INC. · posted Apr 10, 2020 · Open, Classified
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
- Z-1674-2020 · Beckman Coulter Inc. · posted Apr 6, 2020 · Terminated
Due to customer reports it has been determined that the inoculation system contains the incorrect prompt wands. Use of the incorrect wands could result in false -susceptible results, false-resistant results, possible misidentifications, or possible…
- Z-1638-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1639-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1640-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1641-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1642-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1643-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1644-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1645-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1646-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1647-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1648-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1649-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1650-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1651-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1652-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1653-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1654-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1655-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1656-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- BIOMET StageOne, Shoulder Cement Spacer Mold 6 MM, 42 X 18 X 46 MM, Silicone, Sterile, Item 431406.Class IIZ-1657-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- BIOMET StageOne, Shoulder Cement Spacer Mold 8 MM, 46 X 18 X 53 MM, Silicone, Sterile, Item 431408.Class IIZ-1658-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- BIOMET StageOne, Shoulder Cement Spacer Mold 10 MM, 50 X 21 X 57 MM, Silicone, Sterile, Item 431410.Class IIZ-1659-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- BIOMET StageOne, Shoulder Cement Spacer Mold 12 MM, 54 X 21 X 64 MM, Silicone, Sterile, Item 431412.Class IIZ-1660-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- BIOMET StageOne, Shoulder Cement Spacer Mold 14 MM, 58 X 24 X 64 MM, Silicone, Sterile, Item 431414.Class IIZ-1661-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1662-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1663-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1664-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1665-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1666-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1667-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1668-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1669-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1670-2020 · Biomet, Inc. · posted Apr 3, 2020 · Terminated
Potentially insufficient cleaning process or potential inadequate process monitoring for cleaning parameters.
- Z-1630-2020 · Natus Neurology DBA Excel Tech., Ltd. (XLTEK) · posted Apr 1, 2020 · Terminated
The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.
- Z-1631-2020 · Ra Medical Systems, Inc. · posted Apr 1, 2020 · Terminated
At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device to fail to cease firing as long as the foot pedal or handpiece switch is depressed. This malfunction can cause…
- Z-1633-2020 · Physio-Control, Inc. · posted Apr 1, 2020 · Terminated
Automated External Defibrillators may not detect a patient connection due to mechanical wear-through of the contact plating on the therapy connector. The wear-through could result in the device not recognizing a patient is connected which could result in…
- 8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI:… ↗Class IIZ-1635-2020 · Trilliant Surgical, LLC · posted Apr 1, 2020 · Terminated
The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there…
- Z-1729-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
- Z-1730-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
- Z-1731-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
- Z-1732-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
- Z-1733-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
- Z-1734-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
- Z-1735-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
- Z-1736-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
- Z-1737-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
- Z-1738-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
- Z-1739-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
- Z-1740-2020 · Boston Scientific · posted Apr 1, 2020 · Open, Classified
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device