Merlin recallZ-1700-2020 · Abbott · initiated Feb 28, 2020
Class II Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Internal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained.
FDA root cause category: Under Investigation by firm
What the company is doing
On 02/28/2020, the frim sent all US consignees an "IMPORTANT MEDICAL DEVICE CORRECTION" letter via USPS three to five-day ground service. The customer notification informed customer that they had recently deployed a Patient Care System (PCS) programmer software that contained an anomaly that may result in specific programming changes not being stored in the device. The Recalling Firm explained that when using the affected programmer software to make certain programming changes, the changes may appear to have been successfully programmed but the device could remain at previously programmed settings. This issue can occur only if changes to tachycardia therapy Tilt or Waveform (monophasic/biphasic) settings are made and therapy energy value are modified from nominal using the new programmer software. The customer notification also informs customer of Patient Management Recommendations: The Recalling Firm has developed an updated programmer software version which will correct this issue. This solution does not present additional risks to patients and no intervention is needed for patients. The Recalling Firm will update the programmer software in the coming days. For any questions contact your representative or Technical Support at 1-800-722-3774 (U.S.).
Recall details
- Recall number
- Z-1700-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 28, 2020
- Recalling firm
- Abbott
- Quantity
- Software Version 24.6.1 installed into 4733 programmers
- Product code
- NIK
- FDA submissions
- P030035S177, P910023S422, P030054S373
- Distribution
- U.S. AK, Al, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. OUS: None - the firm stated that there are no OUS consignees associated to ICDs that were interrogated from January 22, 2020 to February 6, 2020.
Product description
Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate, program, display data, and test St. Jude MedicalTM implantable devices and leads.
Code information (models, lots, serial numbers)
Software Version: 24.6.1 Model: 3650, GTIN: 05414734500661 Model: 3330, GTIN: 05414734509725 For affected serial number of programmers see Attachment B.1 Programmers For models of ICDs interrogated by affected programmers see Attachment B.2 Devices.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Merlin™ 3650St. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)1 recall1 death380 reports
- Merlin™ 3330St. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)6 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page