Skip to content

Merlin™ 3330Recalls & reported problems

St. Jude Medical, Inc. · PCS 3330 Software Upgrade Kit · Class 3

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
6

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

6 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. May 12, 2026Open, Classified

    Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.

    Z-2405-2026 · Root cause: Software Design Change · FDA record ↗

  2. Nov 21, 2024Open, Classified

    Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.

    Z-0753-2025 · Root cause: Under Investigation by firm · FDA record ↗

  3. Jun 17, 2022Open, Classified

    There is a potential for programmer software (Merlin PCS) and remote monitoring software application (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers.

    Z-1495-2022 · Root cause: Software design · FDA record ↗

  4. Mar 10, 2022Open, Classified

    Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization therapy defibrillator devices, the programmer may continue to execute the Decrement Test instead of terminating the test and restoring the permanent programmed pacing parameters.

    Z-0964-2022 · Root cause: Software design · FDA record ↗

  5. Feb 28, 2020Terminated

    Internal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained.

    Z-1700-2020 · Root cause: Under Investigation by firm · FDA record ↗

  6. Apr 13, 2018Terminated

    Devices inappropriately displayed a low battery indicator determined to be due to memory corruption.

    Z-0808-2020 · Root cause: Process control · FDA record ↗

Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices

All Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.