Peripheral Catheter Insertion Kit recallZ-1946-2020 · Becton Dickinson & Company · initiated Mar 17, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Iincorrectly packaged without the required lidocaine as labeled
FDA root cause category: Packaging change control
What the company is doing
BD issued Urgent Medical Device Recall letter issued on March 17, 2020 via Fed'x or email states reason for recall, health risk and action to take: 1.Examine your inventory of PowerGlide ST" Midline Catheter Maximal Barrier for product code ST018101D, Lot No. REDW0476. Inventory of this kit lot may be used by the clinician through alternate sourcing of lidocaine prior to initiating the procedure. Alternatively, any remaining inventory may be discarded. BD will provide you with a replacement for each kit discarded. 2.Share this recall notification with all users of the product within your facility to ensure they are also aware of this recall. 3.Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement. 4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse If you require further assistance, please contact: BD Contact Contact Information Customer/Technical Support Phone: 800-290-1689 Monday Friday between 8:00 am to 5:00 pm MST in the United States
Recall details
- Recall number
- Z-1946-2020
- Status
- Terminated
- Initiated by the company
- Mar 17, 2020
- Recalling firm
- Becton Dickinson & Company
- Quantity
- 655 kits
- Product code
- PND
- FDA submissions
- K170158
- Distribution
- AZ, CT, GA, IL, LA, MA, MI, MN, ND, NE, NY, OH, SD, TX),
Product description
PowerGlide ST Midline Catheter Maximal Barrier Kit-18 gauge,10 cm length Catalog Number: ST018101D
Code information (models, lots, serial numbers)
Lot Number: REDW0476 20 Exp. Date: 20OCT30 UDI: (01)00801741139390
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Peripheral Catheter Insertion Kit ST018101DBard Access Systems, Inc. · Midline catheter1 recall8 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.